Key result
A single flush-mounted tip thermistor adequately indicated peak electrode-tissue interface temperature during in vivo radiofrequency ablation, underestimating temperature in only 2 of 51 lesions.
Why the study?
Does a single tip thermistor accurately measure peak electrode-tissue interface temperature during radiofrequency catheter ablation in animal models?
Population
In vitro model using 3 porcine right ventricles and in vivo model using 7 mongrel dogs
Comparison
Radiofrequency energy delivery using a standard… vs Different catheter orientations in vitro…
Design
Preclinical
Authors
Loading...
Supports single-thermistor monitoring in animal RF ablation; leaves open human validation before practice change.
Does a single tip thermistor accurately measure peak electrode-tissue interface temperature during radiofrequency catheter ablation in animal models?
A single flush-mounted tip thermistor is adequate for temperature monitoring during radiofrequency ablation to avoid impedance rises when power is regulated to keep temperatures below 90°C.
McRury et al. (1995) studied Radiofrequency Catheter Ablation (n=10). Radiofrequency catheter ablation with multi-thermistor catheter vs. Standard single thermistor tipped electrode was evaluated on Peak electrode-tissue interface temperature accuracy. A single flush-mounted tip thermistor adequately indicated peak electrode-tissue interface temperature during in vivo radiofrequency ablation, underestimating temperature in only 2 of 51 lesions.