Key result
Repositioning stylet-driven leads after failed LBBAP linked to helix fracture and extraction failure in 2 cases.
Why the study?
What are the potential complications of repositioning stylet-driven pacing leads with an extendable helix during left bundle branch area pacing?
What are the potential complications of repositioning stylet-driven pacing leads with an extendable helix during left bundle branch area pacing?
Repositioning stylet-driven leads with an extendable helix during LBBAP carries a risk of helix mechanism fracture and extraction failure.
Answers are grounded in this paper and cite their sources.
Supports caution with stylet-driven lead repositioning in LBBAP; leaves open need for prospective safety data.
Conventional stylet-driven leads with extendable helix can be implanted successfully for left bundle branch area pacing (LBBAP) with a low acute complication rate. We report two cases in which lead repositioning after a first unsuccessful attempt to LBBAP was associated with fracture of the helix rotating mechanism and failure to fully extract the pacing lead.
No takes yet. Share an insight, caveat, or question.
Waroux et al. (2021) studied this question. Repositioning stylet-driven leads with an extendable helix after unsuccessful left bundle branch area pacing caused helix fracture and extraction failure in two cases.