Key result
Analytic model confirms activation-recovery interval bias for positive T waves depends on electrode configuration.
This study identifies the source of measurement bias in activation-recovery intervals from unipolar electrograms with positive T waves and provides recommendations to minimize measurement error.
Electrode configuration adjustments may reduce activation-recovery interval bias in research; leaves open human validation before clinical mapping use.
The activation-recovery interval (ARI) calculated from unipolar electrograms is regularly used as a convenient surrogate measure of local cardiac action potential durations (APD). This method enables important research bridging between computational studies and in vitro and in vivo human studies. The Wyatt method is well established as a theoretically sound method for calculating ARIs; however, some studies have observed that it is prone to a bias error in measurement when applied to positive T waves. This article demonstrates that recent theoretical and computational studies supporting the use of the Wyatt method are likely to have underestimated the extent of this bias in many practical experimental recording scenarios. This work addresses these situations and explains the measurement bias by adapting existing theoretical expressions of the electrogram to represent practical experimental recording configurations. A new analytic expression for the electrogram's local component is derived, which identifies the source of measurement bias for positive T waves. A computer implementation of the new analytic model confirms our hypothesis that the bias is systematically dependent on the electrode configuration. These results provide an aid to electrogram interpretation in general, and this work's outcomes are used to make recommendations on how to minimize measurement error.
No takes yet. Share an insight, caveat, or question.
Western et al. (2014) studied this question. New analytic expression for electrogram vs. Wyatt method was evaluated on Measurement bias in activation-recovery intervals. A new analytic model of unipolar electrograms confirmed that measurement bias in activation-recovery intervals for positive T waves is systematically dependent on the electrode configuration.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: