Why the study?
Does a model-based procedure for automated boundary detection and 3D reconstruction improve stroke volume estimation compared to the conventional bi-plane Simpson method?
Does a model-based procedure for automated boundary detection and 3D reconstruction improve stroke volume estimation compared to the conventional bi-plane Simpson method?
An automated model-based procedure for 3D reconstruction from 2D echocardiograms improves the accuracy of stroke volume estimation compared to the conventional bi-plane Simpson method.
Answers are grounded in this paper and cite their sources.
May refine stroke volume estimates via automated 3D reconstruction; hypothesis-generating and requires prospective validation before practice change.
Describes an automated procedure for boundary detection and 3D reconstruction in echocardiography. Boundary detection is carried out by a combination of knowledge based and dynamic programming procedures. The knowledge based procedure includes a mechanism which controls and evaluates a set of image analysis procedures. The dynamic programming procedure is based on the minimization of a cost function including terms representing echo intensity, edge strength, and boundary continuity in space and time. The 3D reconstruction procedure is based on extracted boundaries from two short-axis and three apical views in conjunction with a geometric model. By replacing conventional bi-plane Simpson method for stroke volume estimation with the authors' model-based scheme, correlation between echocardiographic and dye dilution based estimates increased from 0.69 to 0.88.< <ETX xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">></ETX>
No takes yet. Share an insight, caveat, or question.
Gustavsson et al. (2002) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: