Key result
Biphasic waveform pulses for external defibrillation resulted in lower mean defibrillation thresholds (107 vs 172 J; P=0.04) and shorter resuscitation times compared to monophasic pulses in dogs.
Why the study?
Does external defibrillation with biphasic waveforms improve hemodynamics and reduce defibrillation thresholds compared to monophasic waveforms in a canine model of prolonged ventricular fibrillation?
Population
26 dogs in a model of prolonged ventricular fibrillation
Comparison
External countershocks with biphasic waveforms vs External countershocks with monophasic waveforms
Design
Preclinical, randomized
Authors
Loading...
May guide veterinary resuscitation; leaves open translation to human external defibrillation.
RCT (n=26)
Does external defibrillation with biphasic waveforms improve hemodynamics and reduce defibrillation thresholds compared to monophasic waveforms in a canine model of prolonged ventricular fibrillation?
Absolute Event Rate: 107% vs 172%
p-value: p=0.04
Biphasic waveforms are more efficacious than monophasic waveforms for external defibrillation, resulting in lower defibrillation thresholds and shorter resuscitation times in a canine model of prolonged VF.
Leng et al. (2000) conducted an RCT in Ventricular fibrillation (n=26). Biphasic waveform pulses vs. Monophasic waveform pulses was evaluated on Mean defibrillation threshold after prolonged ventricular fibrillation (p=0.04). Biphasic waveform pulses for external defibrillation resulted in lower mean defibrillation thresholds (107 vs 172 J; P=0.04) and shorter resuscitation times compared to monophasic pulses in dogs.