Why the study?
Does a novel T2 mapping sequence with a 3-parameter fit model improve the accuracy of myocardial T2 values compared to a conventional two-parameter model in healthy subjects and phantoms?
Does a novel T2 mapping sequence with a 3-parameter fit model improve the accuracy of myocardial T2 values compared to a conventional two-parameter model in healthy subjects and phantoms?
A novel T2 mapping sequence with a 3-parameter fit model allows for more accurate and robust measurement of myocardial T2 values independent of T2 prep echo times.
Answers are grounded in this paper and cite their sources.
May support robust T2 mapping in research protocols; leaves open clinical validation and patient outcome impact.
Purpose To develop an improved T 2 prepared (T 2 prep) balanced steady‐state free‐precession (bSSFP) sequence and signal relaxation curve fitting method for myocardial T 2 mapping. Methods Myocardial T 2 mapping is commonly performed by acquisition of multiple T 2 prep bSSFP images and estimating the voxel‐wise T 2 values using a two‐parameter fit for relaxation. However, a two‐parameter fit model does not take into account the effect of imaging pulses in a bSSFP sequence or other imperfections in T 2 prep RF pulses, which may decrease the robustness of T 2 mapping. Therefore, we propose a novel T 2 mapping sequence that incorporates an additional image acquired with saturation preparation, simulating a very long T 2 prep echo time. This enables the robust estimation of T 2 maps using a 3‐parameter fit model, which captures the effect of imaging pulses and other imperfections. Phantom imaging is performed to compare the T 2 maps generated using the proposed 3‐parameter model with the conventional two‐parameter model, as well as a spin echo reference. In vivo imaging is performed on eight healthy subjects to compare the different fitting models. Results Phantom and in vivo data show that the T 2 values generated by the proposed 3‐parameter model fitting do not change with different choices of the T 2 prep echo times, and are not statistically different than the reference values for the phantom ( P = 0.10 with three T 2 prep echoes). The two‐parameter model exhibits dependence on the choice of T 2 prep echo times and are significantly different than the reference values ( P = 0.01 with three T 2 prep echoes). Conclusion The proposed imaging sequence in combination with a three‐parameter model allows accurate measurement of myocardial T 2 values, which is independent of number and duration of T 2 prep echo times. Magn Reson Med 74:93–105, 2015.
No takes yet. Share an insight, caveat, or question.
Akçakaya et al. (2014) studied this question.