Key result
Subcutaneous far field ECG successfully detected R waves with 3% over-detection and 9% under-detection, producing R to R variability results concordant with surface ECG.
Why the study?
Does subcutaneous far field ECG (SFFECG) accurately detect R-waves and R-R variability compared to surface ECG in patients with atrial arrhythmia?
Observational (n=16)
Does subcutaneous far field ECG (SFFECG) accurately detect R-waves and R-R variability compared to surface ECG in patients with atrial arrhythmia?
Answers are grounded in this paper and cite their sources.
Subcutaneous far field ECG is feasible for detecting R-waves and R-R variability, potentially aiding atrial arrhythmia discrimination in implantable devices.
BACKGROUND: Atrial arrhythmia (AA) discrimination remains a technological challenge for implanted cardiac devices. We examined the feasibility of R-wave detection by a subcutaneous far field ECG (SFFECG) and analysis of these signals for R to R variability as an indicator of atrial arrhythmia (AA). METHODS: Surface ECG and SFFECG (from the pacemaker pocket) were recorded in sixteen patients (61.5 +/- 11.4 years) with AA. The SFFECG was recorded with a pacemaker sized four electrode array acutely placed in the pacemaker pocket during implantation. The signals were analyzed to obtain peak-to-peak R wave amplitude and R to R interval variability (indicative of AAs). RESULTS: In sixteen patients R waves were visually discernible in all recordings. The percentage over and under detection for automatic R wave recognition SFFECG was 3 and 9%, respectively. R to R variability analysis using the SFFECG produced results concordant to those using the surface ECG. CONCLUSION: SFFECG might be a helpful adjunct in implantable device systems for detection of R waves and may be used for measurement of R to R variability.
No takes yet. Share an insight, caveat, or question.
Fotuhi et al. (2001) conducted an observational in Atrial arrhythmia (n=16). Subcutaneous far field ECG (SFFECG) vs. Surface ECG was evaluated on R-wave detection and R to R variability. Subcutaneous far field ECG successfully detected R waves with 3% over-detection and 9% under-detection, producing R to R variability results concordant with surface ECG.