Key result
Subcutaneous electrode array electrograms from pacemaker pockets showed sufficient signal-to-noise ratios (average 11 dB vs 27 dB intracardiac) for diagnostic assessment and future device features.
Why the study?
Are subcutaneous electrode array (SEA) electrograms recorded from pacemaker pockets of sufficient amplitude and signal quality for diagnostic assessment?
Observational (n=48)
Are subcutaneous electrode array (SEA) electrograms recorded from pacemaker pockets of sufficient amplitude and signal quality for diagnostic assessment?
Subcutaneous electrode array electrograms from pacemaker pockets provide sufficient signal quality for potential use in electrogram telemetry and diagnostic data storage.
Answers are grounded in this paper and cite their sources.
Supports pocket SEA electrograms for diagnostics; hypothesis-generating for device integration pending prospective validation.
Electrogram signals recorded from typical pacemaker implantation sites may be useful for a variety of pacemaker system functions including pacemaker follow-up, atrial and ventricular sensing (event detection), and triggered electrogram storage. We quantified the electrical characteristics of pacemaker pocket electrograms using a subcutaneous electrode array (SEA) in a population of 48 patients undergoing initial or replacement pacemaker implantation. SEA recorded intrinsic R wave amplitudes measured peak to peak averaged 118 microV and 65 microV for the two recorded SEA electrograms and were significantly different (P < 0.001); paced R wave amplitudes averaged 180 microV and 110 microV. P wave amplitudes averaged 39 microV and 26 microV. No statistically significant difference in amplitudes were observed between acute versus chronic pacemaker pocket or indication for pacing (AV block, sick sinus syndrome). Signal to noise ratios, using R wave amplitude as signal, were lower in the SEA electrogram on average (11 dB) compared to the intracardiac electrogram (27 dB), but sufficient for diagnostic assessment. R wave/P wave ratios for SEA signals were lower than surface and intracardiac values 3.1 and 2.7 compared to a range of 6.2-9.8, indicating a relative enhancement of P waves to R waves in SEA signals. In summary, SEA electrograms are of sufficient amplitude and signal quality (signal to noise ratio) to hold promise for future implantable device features such as electrogram telemetry, enhanced sensing, and diagnostic data storage.
No takes yet. Share an insight, caveat, or question.
Theres et al. (1998) conducted an observational in Pacemaker implantation (n=48). Subcutaneous electrode array (SEA) vs. Intracardiac electrogram was evaluated on Intrinsic R wave amplitudes for the two recorded SEA electrograms (p=<0.001). Subcutaneous electrode array electrograms from pacemaker pockets showed sufficient signal-to-noise ratios (average 11 dB vs 27 dB intracardiac) for diagnostic assessment and future device features.