Evoked endocardial T-wave sensing is highly reliable in pacemaker patients, particularly when using electrodes with a small surface area, porous structure, and atraumatic fixation.
Answers are grounded in this paper and cite their sources.
Supports optimized electrode selection for evoked T-wave sensing; leaves open prospective outcome trials before clinical adoption.
Intracardiac measurements of the evoked endocardial T-wave were performed on 1,500 pacemaker patients. The sensed evoked T-wave amplitude was found to be above 0.75 mV in 94.1 percent of all cases. No correlation was found between the amplitude of the spontaneous R-wave and that of the evoked T-wave. Some electrode characteristics proved to be more favorable than others for evoked endocardial wave sensing, i.e., a surface area of less than 12 mm2, porous surface structure (preferably carbon), and atraumatic fixation.
No takes yet. Share an insight, caveat, or question.
BOUTE et al. (1986) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: