Why the study?
Does ARFI imaging accurately visualize and quantify cardiac radiofrequency ablation lesion dimensions in vitro compared to gross pathology?
Does ARFI imaging accurately visualize and quantify cardiac radiofrequency ablation lesion dimensions in vitro compared to gross pathology?
ARFI imaging accurately visualizes myocardial RFA lesion dimensions to within 2 mm in vitro, highlighting its potential to improve transcatheter cardiac ablation procedures by monitoring lesion formation.
ARFI may aid preclinical ablation lesion assessment; leaves open human validation before clinical use.
INTRODUCTION: Lesion placement and transmurality are critical factors in the success of cardiac transcatheter radiofrequency ablation (RFA) treatments for supraventricular arrhythmias. This study investigated the capabilities of catheter transducer based acoustic radiation force impulse (ARFI) ultrasound imaging for quantifying ablation lesion dimensions. METHODS AND RESULTS: RFA lesions were created in vitro in porcine ventricular myocardium and imaged with an intracardiac ultrasound catheter transducer capable of acquiring spatially registered B-mode and ARFI images. The myocardium was sliced along the imaging plane and photographed. The maximum ARFI-induced displacement images of the lesion were normalized and spatially registered with the photograph by matching the surfaces of the tissue in the B-mode and photographic images. The lesion dimensions determined by a manual segmentation of the photographed lesion based on the visible discoloration of the tissue were compared to automatic segmentations of the ARFI image using 2 different calculated thresholds. ARFI imaging accurately localized and sized the lesions within the myocardium. Differences in the maximum lateral and axial dimensions were statistically below 2 mm and 1 mm, respectively, for the 2 thresholding methods, with mean percent overlap of 68.7 +/- 5.21% and 66.3 +/- 8.4% for the 2 thresholds used. CONCLUSION: ARFI imaging is capable of visualizing myocardial RFA lesion dimensions to within 2 mm in vitro. Visualizing lesions during transcatheter cardiac ablation procedures could improve the success of the treatment by imaging lesion line discontinuity and potentially reducing the required number of ablation lesions and procedure time.
No takes yet. Share an insight, caveat, or question.
Eyerly‐Webb et al. (2009) studied this question.