Key result
The Omron HEM-7130, HEM-7320F, and HEM-7500F automatic blood pressure devices passed all requirements of the ESH-IP 2010 revision for home measurement of BP in adults.
Why the study?
Do the Omron HEM-7130, HEM-7320F, and HEM-7500F automatic blood pressure devices accurately measure blood pressure compared to a standard mercury sphygmomanometer in adults?
Population
33 participants per experiment
Comparison
Omron HEM-7130, HEM-7320F, and HEM-7500F… vs Standard mercury sphygmomanometer
Design
Other
Authors
Loading...
Supports home BP monitoring with these devices; extends validation data but leaves outcome effects open.
Observational (n=99)
Do the Omron HEM-7130, HEM-7320F, and HEM-7500F automatic blood pressure devices accurately measure blood pressure compared to a standard mercury sphygmomanometer in adults?
The Omron HEM-7130, HEM-7320F, and HEM-7500F automatic blood pressure devices meet the ESH-IP 2010 validation requirements and are recommended for home blood pressure monitoring.
Takahashi et al. (2014) conducted an observational in blood pressure monitoring (n=99). Omron HEM-7130, HEM-7320F, and HEM-7500F vs. standard mercury sphygmomanometer was evaluated on mean differences between the device and the mercury readings for systolic and diastolic BP. The Omron HEM-7130, HEM-7320F, and HEM-7500F automatic blood pressure devices passed all requirements of the ESH-IP 2010 revision for home measurement of BP in adults.