Key result
The standard 0.5 mV left atrial scar cutoff varies significantly by catheter design.
Why the study?
Bipolar voltage electrograms for left atrial substrate characterization depend on catheter design and electrode configuration.
Observational (n=25)
Bipolar voltage amplitude cutoffs for left atrial scar and healthy tissue vary significantly depending on catheter design, necessitating catheter-specific adapted cutoffs.
Catheter-specific voltage cutoffs may refine left atrial scar mapping; hypothesis-generating pending prospective outcome validation.
BACKGROUND: Bipolar voltage (BV) electrograms for left atrial (LA) substrate characterization depend on catheter design and electrode configuration. AIMS: The aim of the study was to investigate the relationship between the BV amplitude (BVA) using four catheters with different electrode design and to identify their specific LA cutoffs for scar and healthy tissue. METHODS AND RESULTS: Consecutive high-resolution electroanatomic mapping was performed using a multipolar-minielectrode Orion catheter (Orion-map), a duo-decapolar circular mapping catheter (Lasso-map), and an irrigated focal ablation catheter with minielectrodes (Mifi-map). Virtual remapping using the Mifi-map was performed with a 4.5 mm tip-size electrode configuration (Nav-map). BVAs were compared in voxels of 3 × 3 × 3 mm3. The equivalent BVA cutoff for every catheter was calculated for established reference cutoff values of 0.1, 0.2, 0.5, 1.0, and 1.5 mV. We analyzed 25 patients (72% men, age 68 ± 15 years). For scar tissue, a 0.5 mV cutoff using the Nav corresponds to a lower cutoff of 0.35 mV for the Orion and of 0.48 mV for the Lasso. Accordingly, a 0.2 mV cutoff corresponds to a cutoff of 0.09 mV for the Orion and of 0.14 mV for the Lasso. For healthy tissue cutoff at 1.5 mV, a larger BVA cutoff for the small electrodes of the Orion and the Lasso was determined of 1.68 and 2.21 mV, respectively. CONCLUSION: When measuring LA BVA, significant differences were seen between focal, multielectrode, and minielectrode catheters. Adapted cutoffs for scar and healthy tissue are required for different catheters.
No takes yet. Share an insight, caveat, or question.
Knecht et al. (2023) conducted an observational in Left atrial substrate characterization (n=25). Multipolar, focal and mini-electrode catheters was evaluated on Equivalent bipolar voltage amplitude cutoffs for scar and healthy tissue. Left atrial bipolar voltage amplitude cutoffs varied significantly by catheter design, with a 0.5 mV focal scar cutoff corresponding to 0.35 mV for Orion and 0.48 mV for Lasso catheters.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: