Key result
Temperature-guided radiofrequency ablation in a beating porcine heart showed a positive correlation between lesion size and peak temperature in bipolar mode.
Why the study?
Does temperature-guided radiofrequency ablation produce predictable lesion sizes in an open-chest pig model?
Does temperature-guided radiofrequency ablation produce predictable lesion sizes in an open-chest pig model?
In a beating heart pig model, it is difficult to predict RF ablation lesion size from temperature or energy, particularly in unipolar configuration.
Answers are grounded in this paper and cite their sources.
May support temperature guidance in bipolar RF ablation; leaves open predictability in unipolar clinical use.
The safety and feasibility of temperature-guided radiofrequency (RF) ablation of ventricular myocardium were assessed in an open-chest animal model. RF ablation was performed in the ventricles of 19 pigs using ablation catheters with a tip electrode of 2 or 4 mm length. The energy was delivered in a bipolar (2 mm tip electrode only) or unipolar configuration. Set temperature ranged from 60 to 90 degrees C. Pulse duration was 30 s. Histological examination showed coagulation necrosis with a haemorrhagic zone. However, lesions with a deep cleavage were found after five unipolar (2 mm tip electrode) energy applications. No such lesions were found after unipolar energy applications with a 4 mm tip electrode. During or shortly after ablation, premature ventricular beats and non-sustained VT were frequently observed and in some instances ventricular fibrillation. In the bipolar mode we found a positive correlation between lesion area/volume and peak temperature with a correlation coefficient of 0.48 and 0.56, respectively (P < 0.05). However no correlation was found between lesion size and applied energy. In the unipolar configuration there was no correlation between lesion size and these parameters. Median depth and area were: bipolar: 4.0 mm/23.5 mm2, unipolar (2 mm tip electrode): 3.5 mm/12.2 mm2, unipolar (4 mm tip electrode): 4.0 mm/15.7 mm2. We conclude that in a beating heart it is difficult to predict lesion size from temperature or energy.
No takes yet. Share an insight, caveat, or question.
Kongsgaard et al. (1993) studied Ventricular myocardium ablation (n=19). Temperature-guided radiofrequency ablation was evaluated on Correlation between lesion area/volume and peak temperature (correlation coefficient 0.48 and 0.56, p=<0.05). Temperature-guided radiofrequency ablation in a beating porcine heart showed a positive correlation between lesion size and peak temperature in bipolar mode.
Synapse has enriched 10 related papers. Consider them for comparative context: