Key result
MarioHeart in-vitro model generates physiological pulsatile flow at 5.3 L/min and replicates mechanical valve thrombosis.
Why the study?
Development of more hemocompatible mechanical heart valves and new anticoagulants is impeded by a lack of adequate in-vitro models.
The MarioHeart in-vitro model successfully emulates physiological pulsatile flow and demonstrates feasibility for testing the thrombogenicity of mechanical heart valves.
Answers are grounded in this paper and cite their sources.
May enable preclinical thrombogenicity testing of mechanical valves; hypothesis-generating and leaves open clinical translation.
Mechanical heart valve (MHV) prostheses present a risk of thromboembolic complications despite antithrombotic therapy. Further steps in the development of more hemocompatible MHVs and new anticoagulants are impeded due to the lack of adequate in-vitro models. With the development of a novel in-vitro model (MarioHeart), a pulsatile flow similar to the arterial circulation is emulated. The MarioHeart design owns unique features as 1) a single MHV within a torus with low surface/volume ratio, 2) a closed loop system, and 3) a dedicated external control system driving the oscillating rotational motion of the torus. For verification purposes, a blood analog fluid seeded with particles was used to assess fluid velocity and flow rate using a speckle tracking method on high-speed video recordings of the rotating model. The flow rate resembled the physiological flow rate in the aortic root, in both shape and amplitude. Additional in-vitro runs with porcine blood showed thrombi on the MHV associated with the suture ring, which is similar to the in-vivo situation. MarioHeart is a simple design which induces well-defined fluid dynamics resulting in physiologically nonturbulent flow without stasis of the blood. MarioHeart seems suitable for testing the thrombogenicity of MHVs and the potential of new anticoagulants.
No takes yet. Share an insight, caveat, or question.
Devos et al. (2023) studied Mechanical heart valve thrombosis. MarioHeart in-vitro flow model was evaluated on Fluid velocity, flow rate, and thrombus formation. The MarioHeart in-vitro flow model generated physiologically similar pulsatile flow with a mean flow of 5.3 L/min and demonstrated thrombus formation on mechanical heart valves using porcine blood.