Key result
A novel in vitro pulsatile flow facility successfully simulated physiological conditions to evaluate the hemodynamics and thrombogenicity of a 23 mm St. Jude Medical mechanical heart valve.
The developed in vitro pulsatile flow facility is capable of simultaneously characterizing blood damage and fluid dynamics for prosthetic heart valves.
No takes yet. Share an insight, caveat, or question.
Novel in vitro facility enables simultaneous valve testing; leaves open validation against clinical outcomes before guiding practice.
Arjunon et al. (2015) studied Prosthetic heart valve thrombogenicity. Pulsatile flow facility test-platform was evaluated on Blood damage (TAT, PF1.2, hemolysis) and fluid dynamics. A novel in vitro pulsatile flow facility successfully simulated physiological conditions to evaluate the hemodynamics and thrombogenicity of a 23 mm St. Jude Medical mechanical heart valve.