Key result
Point-of-care AI-ECG algorithm fails to detect asymptomatic SARS-CoV-2 infection.
Why the study?
To evaluate the ability of a neural network to identify SARS-CoV-2 infection using point-of-care electrocardiography obtained with a portable device.
Does a point-of-care AI-ECG accurately detect asymptomatic SARS-CoV-2 infection in adults?
Observational (n=2,827)
Does a point-of-care AI-ECG accurately detect asymptomatic SARS-CoV-2 infection in adults?
A point-of-care AI-ECG failed to reliably identify asymptomatic SARS-CoV-2 infection, highlighting the necessity of prospective validation of AI tools in target populations.
Answers are grounded in this paper and cite their sources.
Does not support point-of-care AI-ECG for asymptomatic SARS-CoV-2 screening; leaves open the need for prospective validation of AI tools.
Objective: To evaluate the ability of a neural network to identify severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) infection using point-of-care electrocardiography obtained with a portable device. Patient and Methods: We enrolled 2827 patients in a prospective observational study, from December 10, 2020, through June 4, 2021, to determine the accuracy of a point-of-care, handheld, smartphone-compatible, artificial intelligence-enabled electrocardiography (ECG) (POC AI-ECG) in detecting asymptomatic SARS-CoV-2 infection using a modified version of an existing deep learning model framework trained on 12-lead ECG data. Results: Study participants were 48% (n=1067) female, 79% (n=1749) White, and 7% (n=153) endorsed previous COVID-19 infection. We found the POC AI-ECG algorithm was ineffective for detecting asymptomatic SARS-CoV-2 infection (area under curve, 0.56; 95% CI, 0.46-0.66), failing to adequately discriminate between ECGs performed among participants who tested positive compared to those who tested negative. Conclusion: Contrary to the prior 12-lead ECG study, a POC AI-ECG failed to reliably identify asymptomatic SARS-CoV-2 infection among adults. This study underscores the importance of prospective testing, assuring similar populations, and using similar signals or data when developing AI-ECG tools. Trial registration: clinicaltrials.gov Identifier: NCT04725097.
No takes yet. Share an insight, caveat, or question.
Adedinsewo et al. (2023) conducted an observational in Asymptomatic SARS-CoV-2 infection (n=2,827). Point-of-care artificial intelligence-enabled electrocardiography (POC AI-ECG) vs. Standard SARS-CoV-2 testing was evaluated on Detection of asymptomatic SARS-CoV-2 infection (AUC 0.56, 95% CI 0.46-0.66). A point-of-care artificial intelligence-enabled electrocardiography algorithm was ineffective for detecting asymptomatic SARS-CoV-2 infection.
Synapse has enriched 10 related papers. Consider them for comparative context: