Key result
Direct self-expanding valve implantation yields similar 1-year mortality compared to pre-dilatation.
Why the study?
Intermediate-term data from randomized studies evaluating the safety and efficacy of direct transcatheter aortic valve implantation are lacking.
Does direct implantation of a self-expanding valve without pre-dilatation improve clinical and echocardiographic outcomes at 1 year in patients with severe aortic stenosis compared to pre-dilatation?
RCT (n=171)
Yes
Does direct implantation of a self-expanding valve without pre-dilatation improve clinical and echocardiographic outcomes at 1 year in patients with severe aortic stenosis compared to pre-dilatation?
Direct implantation of a self-expanding valve without pre-dilatation yields similar 1-year clinical and echocardiographic outcomes compared to implantation with pre-dilatation.
Answers are grounded in this paper and cite their sources.
Supports omitting pre-dilatation before self-expanding TAVR; confirms similar 1-year mortality and hemodynamics in this RCT.
Objective To present 1 year clinical and echocardiographic outcomes of the randomized DIRECT (Pre‐dilatation in Transcatheter Aortic Valve Implantation Trial) trial. Background Intermediate‐term data from randomized studies investigating the safety and efficacy of direct implantation are lacking. Methods DIRECT trial randomized 171 consecutive patients with severe aortic stenosis at four tertiary centers to undergo TAVI with the use of self‐expanding prostheses with (pre‐BAV) or without pre‐dilatation (no‐BAV). The primary endpoint was device success according to the VARC‐2 criteria. All patients underwent a clinical and echocardiographic follow‐up at 1 year. All‐cause and cardiac mortality, stroke, heart failure hospitalization, and new pacemaker implantation were recorded. Results At 1 year, four deaths were recorded in pre‐BAV group (4.7%) and three deaths in no‐BAV group (3.5%). There was no difference in Kaplan–Meier plots between the two groups in all‐cause mortality at 1 year (log‐rank p = .72). Similarly, there was no difference in the incidence of permanent pacemaker implantation between the two groups at 1 year (27/67–40.3% in no‐BAV group versus 20/69–29% in pre‐BAV group, log‐rank p = .24). There was no significant difference between pre‐BAV and no BAV group in aortic valve area (1.84 ± 0.39 cm 2 vs. 1.85 ± 0.44 cm 2 , p = .90), mean aortic valve gradient (8.36 ± 5.04 vs. 8.00 ± 4.04 mmHg, p = .65) and moderate or severe paravalvular regurgitation (5–6.6 vs. 4–5.7%, respectively) at 1 year. The same applied independently from the performance of post‐dilatation at baseline. Conclusions Direct, without pre‐dilatation, implantation of a self‐expanding valve has no impact on one‐year clinical and echocardiographic outcomes, independently also from the baseline performance of post‐dilatation.
No takes yet. Share an insight, caveat, or question.
Benetos et al. (2020) conducted an RCT in severe aortic stenosis (n=171). Self-expanding prostheses without pre-dilatation (no-BAV) vs. Self-expanding prostheses with pre-dilatation (pre-BAV) was evaluated on device success according to the VARC-2 criteria. Direct implantation of a self-expanding valve without pre-dilatation resulted in similar 1-year all-cause mortality compared to pre-dilatation (3.5% vs 4.7%, p=0.72).
Synapse has enriched 10 related papers. Consider them for comparative context: