Key result
PTFE felt pledgets demonstrated consistently higher CT attenuation (420±26 HU) than surrounding contrast-enhanced blood (288±41 HU), aiding in their differentiation from paravalvular leakage.
Why the study?
Does CT attenuation measurement differentiate PTFE felt pledgets from contrast-enhanced blood (paravalvular leakage) in patients with prosthetic heart valves?
Observational (n=19)
Does CT attenuation measurement differentiate PTFE felt pledgets from contrast-enhanced blood (paravalvular leakage) in patients with prosthetic heart valves?
PTFE felt pledgets have consistently higher CT attenuation than contrast-enhanced blood, allowing for their differentiation from paravalvular leakage on multislice CT.
Answers are grounded in this paper and cite their sources.
May reduce false-positive paravalvular leak calls on CT; hypothesis-generating for prosthetic valve protocols.
OBJECTIVES: Sutures with polytetrafluorethylene (PTFE) felt pledgets are commonly used in prosthetic heart valve (PHV) implantation. Paravalvular leakage can be difficult to distinguish from PTFE felt pledgets on multislice CT because both present as hyperdense structures. We assessed whether pledgets can be discriminated from contrast-enhanced solutions (blood/saline) on CT images based on attenuation difference in an ex vivo experiment and under in vivo conditions. METHODS: PTFE felt pledgets were sutured to the suture ring of a mechanical PHV and porcine aortic annulus, and immersed and scanned in four different contrast-enhanced (Ultravist®; 300 mg jopromide ml(-1)) saline concentrations (10.0, 12.0, 13.6 and 15.0 mg ml(-1)). Scanning was performed on a 256-slice scanner with eight different scan protocols with various tube voltage (100 kV, 120 kV) and tube current (400 mAs, 600 mAs, 800 mAs, 1000 mAs) settings. Attenuation of the pledgets and surrounding contrast-enhanced saline were measured. Additionally, the attenuation of pledgets and contrast-enhanced blood was measured on electrocardiography (ECG)-gated CTA scans of 19 patients with 22 PHVs. RESULTS: Ex vivo CT attenuation differences between the pledgets and contrast-enhanced solutions were larger by using higher tube voltages. CT attenuation values of the pledgets were higher than contrast-enhanced blood in patients: 420±26 Hounsfield units (mean±SD, range 383-494) and 288±41 Hounsfield units (range 202-367), respectively. CONCLUSIONS: PTFE felt pledgets have consistently higher attenuation than surrounding contrast-enhanced blood. CT attenuation measurements therefore may help to differentiate pledgets from paravalvular leakage, and detect paravalvular leakage in patients with suspected PHV dysfunction.
No takes yet. Share an insight, caveat, or question.
Habets et al. (2012) conducted an observational in Prosthetic heart valve (n=19). CT attenuation measurement of PTFE felt pledgets vs. Contrast-enhanced blood was evaluated on CT attenuation of pledgets versus contrast-enhanced blood. PTFE felt pledgets demonstrated consistently higher CT attenuation (420±26 HU) than surrounding contrast-enhanced blood (288±41 HU), aiding in their differentiation from paravalvular leakage.