Key result
Reprogramming a cardiac resynchronization therapy pacemaker with a V-V offset caused inappropriate alarms in a wearable cardioverter-defibrillator due to double counting of the ventricular electrogram.
Why the study?
To report a previously undescribed interaction and alarm from a wearable cardioverter-defibrillator following cardiac resynchronization therapy pacemaker reprogramming with a V-V offset.
Case Report (n=1)
No
Reprogramming a CRT device with a V-V offset in a patient wearing a WCD can cause inappropriate arrhythmia detection and alarms due to double counting and morphology changes.
Answers are grounded in this paper and cite their sources.
Reprogramming CRT with V-V offset may trigger inappropriate WCD alarms; this case leaves open the incidence and mitigation of device interactions.
We describe the case of a patient who received cardiac resynchronization pacemaker therapy (CRT-P) incorporating bridge protection via a wearable cardioverter-defibrillator (WCD) while awaiting final evaluation for permanent placement of an implantable CRT device with defibrillation capabilities. In an attempt to mitigate symptoms of heart failure, reprogramming of the CRT-P with a V-V offset caused an unusual interaction and alarm from the WCD. This case highlights a unique interaction not previously reported and is clinically relevant to patients awaiting final evaluation for implantable defibrillator system placement.
No takes yet. Share an insight, caveat, or question.
Tobb et al. (2020) conducted a case report in Cardiomyopathy with heart failure (n=1). CRT-P reprogramming with V-V offset vs. Simultaneous mode was evaluated on Inappropriate WCD alarms. Reprogramming a cardiac resynchronization therapy pacemaker with a V-V offset caused inappropriate alarms in a wearable cardioverter-defibrillator due to double counting of the ventricular electrogram.
Synapse has enriched 10 related papers. Consider them for comparative context: