Key result
Oscillometric blood pressure monitors had measurement errors highly dependent on patient characteristics (accounting for 36.6% of systolic variation), unlike microphonic monitors (1.2%).
Why the study?
Do patient characteristics affect the accuracy of oscillometric and microphonic ambulatory blood pressure monitors compared to random-zero sphygmomanometry?
Population
118 subjects
Comparison
Blood pressure measurement using oscillometric… vs Simultaneous readings by two pairs of trained…
Design
Cross-sectional, blinded observers
Authors
Loading...
May compromise oscillometric monitor reliability in select patients; leaves open whether microphonic devices warrant preference in validation studies.
Cross-Sectional (n=118)
Blinded observers
Do patient characteristics affect the accuracy of oscillometric and microphonic ambulatory blood pressure monitors compared to random-zero sphygmomanometry?
Oscillometric ambulatory blood pressure monitors are significantly more susceptible to measurement errors driven by patient characteristics such as age and pulse pressure compared to microphonic monitors.
Pannarale et al. (1993) reported a cross-sectional. Oscillometric (SpaceLabs 90207) and microphonic (TM2420) ambulatory blood pressure monitors vs. Random-zero sphygmomanometry by trained observers was evaluated on Measurement error (difference between monitor and trained observers' readings). Oscillometric blood pressure monitors had measurement errors highly dependent on patient characteristics (accounting for 36.6% of systolic variation), unlike microphonic monitors (1.2%).