October 1, 1993HypertensionOpen Access

Bias and variability in blood pressure measurement with ambulatory recorders.

View Full Paper
Ask AI
Bookmark
Share

Key result

Oscillometric blood pressure monitors had measurement errors highly dependent on patient characteristics (accounting for 36.6% of systolic variation), unlike microphonic monitors (1.2%).

Why the study?

Do patient characteristics affect the accuracy of oscillometric and microphonic ambulatory blood pressure monitors compared to random-zero sphygmomanometry?

Population

118 subjects

Comparison

Blood pressure measurement using oscillometric… vs Simultaneous readings by two pairs of trained…

Design

Cross-sectional, blinded observers

Authors

GPGiuseppe PannaraleSapienza University of RomeGBGwyn BebbAmgen (United States)SCSusan J. ClarkUniversity of Sheffield

Discussion

Loading...

Member takes

Implication

May compromise oscillometric monitor reliability in select patients; leaves open whether microphonic devices warrant preference in validation studies.

Key Points

  • The aim was to explore how patient characteristics affect the accuracy of two types of blood pressure monitors.
  • Compared records from two oscillometric and two microphonic monitors with readings from trained observers using random-zero sphygmomanometry.
  • Evaluated 118 subjects aged 17 to 94 with various blood pressure levels.
  • Assessed correlations between measurement errors and patient characteristics using multiple regression.
  • The oscillometric monitor showed an average systolic error of 2.8 mm Hg and diastolic error of 3.9 mm Hg.
  • For oscillometric measurements, patient characteristics accounted for 36.6% of systolic error variation and 34.7% of diastolic error variation.
  • The microphonic monitor had minimal correlation with patient characteristics, accounting for only 1.2% of systolic error.

Study Design

Type

Cross-Sectional (n=118)

Blinding

Blinded observers

Structured PICO

Do patient characteristics affect the accuracy of oscillometric and microphonic ambulatory blood pressure monitors compared to random-zero sphygmomanometry?

P
Population
118 subjects (53 men and 65 women, aged 17 to 94 years; systolic pressure, 89 to 211 mm Hg; diastolic, 44 to 116 mm Hg)
I
Intervention
Blood pressure measurement using oscillometric (SpaceLabs 90207) and microphonic (TM2420 version 7) ambulatory monitors
C
Comparator
Simultaneous readings by two pairs of trained, blinded observers using random-zero sphygmomanometry
O
Outcome
Measurement error (difference between monitor and trained observers' readings) and its correlation with patient characteristicssurrogate

Oscillometric ambulatory blood pressure monitors are significantly more susceptible to measurement errors driven by patient characteristics such as age and pulse pressure compared to microphonic monitors.

Cite This Study

Pannarale et al. (1993) reported a cross-sectional. Oscillometric (SpaceLabs 90207) and microphonic (TM2420) ambulatory blood pressure monitors vs. Random-zero sphygmomanometry by trained observers was evaluated on Measurement error (difference between monitor and trained observers' readings). Oscillometric blood pressure monitors had measurement errors highly dependent on patient characteristics (accounting for 36.6% of systolic variation), unlike microphonic monitors (1.2%).

synapsesocial.com/papers/6a157f71814bf8ec9a4eacbehttps://doi.org/10.1161/01.hyp.22.4.591
View Full Paper
Ask AI
Bookmark
Share

Also Consider

Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context:

  1. 1Bench and ambulatory field evaluation of the A & D TM-2420 automated sphygmomanometer1990 · 19 citations
  2. 2A Comparison of Two Ambulatory Blood Pressure Monitors, the Del Mar Avionics Pressurometer IV and the Spacelabs 902021989 · 34 citations
  3. 3The British Hypertension Society protocol for the evaluation of automated and semi-automated blood pressure measuring devices with special reference to ambulatory systems1990 · 778 citations
  4. 4Accuracy of the SpaceLabs 90207 determined by the British Hypertension Society Protocol1991 · 453 citations
  5. 5Effect of Arterial Compliance on Systolic Blood Pressure and Cardiac Function1982 · 21 citations