Key result
OptiBP/CamBP cuffless BP showed low mean bias vs reference BP in nonpregnant adults with high heterogeneity.
Why the study?
Does smartphone-based cuffless blood pressure estimation using the OptiBP/CamBP platform accurately agree with validated auscultatory or oscillometric reference measurements in nonpregnant adult and pregnant cohorts?
Meta-Analysis
Does smartphone-based cuffless blood pressure estimation using the OptiBP/CamBP platform accurately agree with validated auscultatory or oscillometric reference measurements in nonpregnant adult and pregnant cohorts?
Answers are grounded in this paper and cite their sources.
The OptiBP/CamBP smartphone-based cuffless blood pressure platform demonstrates low average systematic bias compared to reference standards, though substantial heterogeneity and limited data in pregnancy prevent universal clinical interchangeability.
Background: Smartphone-based cuffless blood pressure (BP) estimation could broaden scalable BP screening and monitoring, but clinical adoption requires rigorous agreement with standard methods. Validation studies of the OptiBP/CamBP platform have shown promising but heterogeneous results. We synthesized the available evidence while accounting for repeated paired measurements and clinically distinct cohorts. Methods: We performed a PRISMA-concordant systematic review and random-effects meta-analysis of OptiBP/CamBP validation studies comparing smartphone-derived BP estimates with validated auscultatory or oscillometric reference measurements. Primary outcomes were systolic BP (SBP) and diastolic BP (DBP) mean bias, defined as OptiBP minus reference BP (mmHg). To reduce overprecision from repeated within-participant measurements, participant-level analyzed sample size and subject-level precision metrics were prioritized. Random-effects models used restricted maximum likelihood with Hartung-Knapp inference. Nonpregnant adult and pregnancy cohorts were analyzed separately. Results: Five studies comprising 9 analyzable cohorts were included. In nonpregnant adults, pooled SBP mean bias was 0.77 mmHg (95% CI -2.20 to 3.74; I 2 =79.1%) and pooled DBP mean bias was 0.61 mmHg (95% CI -2.25 to 3.47; I 2 =92.1%), indicating no large systematic directional bias but substantial heterogeneity. In pregnancy cohorts, pooled SBP mean bias was -0.52 mmHg (95% CI -14.93 to 13.89; I 2 =83.8%) and pooled DBP mean bias was 0.39 mmHg (95% CI -5.71 to 6.49; I 2 =39.9%), reflecting imprecision due to limited evidence. In sensitivity analyses combining all article-level strata, pooled SBP and DBP mean bias remained near zero (0.28 mmHg and 0.57 mmHg, respectively). Leave-one-out analyses identified the Festo adult cohort as the main driver of SBP heterogeneity and the Caillat cohort as the main driver of DBP heterogeneity. Conclusions: Across available validation studies, OptiBP/CamBP demonstrated low average systematic bias relative to reference BP measurement for both SBP and DBP. However, substantial between-study heterogeneity and limited pregnancy-specific evidence preclude claims of universal interchangeability across settings. These findings support cautious optimism and underscore the need for larger, independently replicated, methodologically standardized validation studies in diverse and clinically challenging populations.
No takes yet. Share an insight, caveat, or question.
Daneii et al. (2026) conducted a meta-analysis in Blood pressure estimation and monitoring. OptiBP/CamBP smartphone-based cuffless blood pressure estimation vs. Validated auscultatory or oscillometric reference blood pressure measurements was evaluated on Systolic blood pressure (SBP) and diastolic blood pressure (DBP) mean bias (OptiBP minus reference BP in mmHg) (MD 0.77 mmHg (nonpregnant adult SBP); MD 0.61 mmHg (nonpregnant adult DBP), 95% CI -2.20 to 3.74 (SBP); -2.25 to 3.47 (DBP)). OptiBP/CamBP cuffless BP showed low mean bias vs reference BP in nonpregnant adults with high heterogeneity.