Key result
The AlwaysBP smartphone software estimated systolic and diastolic blood pressure with mean differences of 1.2 ± 7.1 mmHg and -1.0 ± 6.0 mmHg, respectively, compared to standard cuff measurements.
Why the study?
Although several mobile applications measuring blood pressure have been developed, data regarding their accuracy are limited.
Does the AlwaysBP smartphone software accurately measure blood pressure compared to standard cuff-based manual measurement in adults?
Observational (n=85)
No
Does the AlwaysBP smartphone software accurately measure blood pressure compared to standard cuff-based manual measurement in adults?
The AlwaysBP smartphone application demonstrated acceptable accuracy for measuring systolic and diastolic blood pressure compared to standard manual cuff-based methods.
Answers are grounded in this paper and cite their sources.
Supports smartphone BP app evaluation; leaves open prospective validation before clinical adoption.
BACKGROUND Even though several mobile applications that can measure blood pressure (BP) have been developed, the data about the accuracy of those applications are limited. OBJECTIVE This study assessed the accuracy of AlwaysBP (test) in BP measurement compared with the standard cuff-based manual method of brachial BP measurement (reference). METHODS AlwaysBP is a smartphone software that estimates systolic BP (SBP) and diastolic BP (DBP) based on the pulse transit time (PTT). The PTT was calculated from the finger photoplethysmogram using the camera and from seismocardiogram using the inertial measurement unit sensor of a commercially available smartphone. After calculating PTT, SBP and DBP were estimated from the Bramwell-Hill and Moens-Korteweg equations. A calibration process was carried out three times for each subject to determine the input parameters of the equation. The study was conducted from March to August 2021 at Chungnam National University Sejong Hospital with 85 participants between the ages of 19 and 70 years who met specific conditions. The primary analysis aimed to evaluate the accuracy of the test method compared with the reference method in the entire study population. The secondary analysis was performed to confirm the stability of the test method for up to 4 weeks in 15 subjects. At enrollment, the gender, arm circumference, and BP distribution were considered according to the current guidelines. RESULTS Among the study subjects, 45 (51.7%) individuals were male and the average age was 35.6 ± 10.4 years. Hypertension was diagnosed in 14 subjects (16.1%) prior to the study. The mean test and reference SBP were 120.0 ± 18.8 and 118.7 ± 20.2 mmHg, respectively, and the difference was 1.2 ± 7.1 mmHg. The absolute difference between the test and baseline SBP was <5 mmHg in 57.5%, <10 mmHg in 84.3%, and <15 mmHg in 94.6% of the subjects. The mean test and reference DBP were 80.1 ± 12.6 and 81.1 ± 14.4 mmHg, respectively, and the difference was −1.0 ± 6.0 mmHg. The absolute difference between the test and the reference DBP was <5 mmHg in 75.5%, <10 mmHg in 93.9%, and <15 mmHg in 97.3% of the subjects. The secondary analysis showed that after 4 weeks, the differences between SBP and DBP were 0.1 ± 8.8 and −2.4 ± 7.6 mmHg, respectively. CONCLUSIONS The AlwaysBP software exhibited acceptable accuracy regarding SBP and DBP measurement compared with the standard BP measurement method according to AAMI/ESH/ISO protocol criteria. However, further validation study is required with a specific validation protocol designed for cuffless BP measuring devices to assess clinical accuracy. This technology can be easily applied in everyday life and may improve the general population’s awareness of hypertension, thus helping to control it. CLINICALTRIAL
No takes yet. Share an insight, caveat, or question.
Yoon et al. (2022) conducted an observational in Blood pressure measurement (n=85). AlwaysBP smartphone software vs. Standard cuff-based manual method of brachial BP measurement was evaluated on Accuracy of systolic blood pressure (SBP) measurement compared with the reference method (MD 1.2). The AlwaysBP smartphone software estimated systolic and diastolic blood pressure with mean differences of 1.2 ± 7.1 mmHg and -1.0 ± 6.0 mmHg, respectively, compared to standard cuff measurements.