Why the study?
This study evaluated the accuracy and stability of AlwaysBP, a smartphone software using pulse transit time to estimate blood pressure, compared with standard cuff-based manual measurement.
Does AlwaysBP smartphone software accurately measure blood pressure compared to standard cuff-based manual measurement in individuals without known cardiovascular disease?
Does AlwaysBP smartphone software accurately measure blood pressure compared to standard cuff-based manual measurement in individuals without known cardiovascular disease?
The AlwaysBP smartphone software demonstrated acceptable accuracy for measuring systolic and diastolic blood pressure compared to standard cuff-based methods, offering a potential tool for improving hypertension awareness.
Answers are grounded in this paper and cite their sources.
Smartphone BP apps need independent validation before any clinical use; leaves open reliability in real-world settings.
Background This study assessed the accuracy of AlwaysBP (test) in blood pressure (BP) measurement compared with the standard cuff-based manual method of brachial BP measurement (reference). AlwaysBP is a smartphone software that estimates systolic BP (SBP) and diastolic BP (DBP) based on the pulse transit time (PTT). The PTT was calculated from the finger photoplethysmogram using the camera and from seismocardiogram using the inertial measurement unit sensor of a commercially available smartphone. Methods A total of 87 participants without known cardiovascular disease were enrolled in the study. The primary analysis aimed to evaluate the accuracy of the test method compared with the reference method in the entire study population. The secondary analysis was performed to confirm the stability of the test method for up to 4 weeks in 15 subjects. At enrollment, the gender, arm circumference, and BP distribution were considered according to the current guidelines. Results Among the study subjects, 45 (51.7%) individuals were male and the average age was 35.6 ± 10.4 years. Hypertension was diagnosed in 14 subjects (16.1%) prior to the study. The mean test and reference SBP were 120.0 ± 18.8 and 118.7 ± 20.2 mmHg, respectively, and the difference was 1.2 ± 7.1 mmHg ( r = 0.935, P < 0.001). The mean test and reference DBP were 80.1 ± 12.6 and 81.1 ± 14.4 mmHg, respectively, and the difference was −1.0 ± 6.0 mmHg ( r = 0.911, P < 0.001). At the 4-week time point, the differences between SBP and DBP were 0.1 ± 8.8 and −2.4 ± 7.6 mmHg. Conclusions The AlwaysBP software exhibited acceptable accuracy regarding SBP and DBP measurement compared with the standard BP measurement method. This technology can be easily applied in everyday life and may improve the general population’s awareness of hypertension, thus helping to control it.
No takes yet. Share an insight, caveat, or question.
Yoon et al. (2022) studied this question.