Optimized ex vivo swine myocardial models achieve similar ablation depths to in vivo canine preparations.
Why the study?
Reliable ex vivo cardiac ablation models can increase catheter testing throughput while minimizing animal usage, motivating the development and parameter optimization of a physiologically relevant ex vivo swine model displaying minimal variability and high repeatability.
Answers are grounded in this paper and cite their sources.
An optimized ex vivo swine model of cardiac ablation produces highly reproducible lesion characteristics comparable to the standard in vivo canine thigh preparation model, potentially reducing animal usage in catheter testing.
INTRODUCTION: Reliable ex vivo cardiac ablation models have the potential to increase catheter testing throughput while minimizing animal usage. The goal of this work was to develop a physiologically relevant ex vivo swine model of cardiac ablation displaying minimal variability and high repeatability and identify and optimize key parameters involved in ablation outcomes. METHODS AND RESULTS: A root cause analysis was conducted to identify variables affecting ablation outcomes. Parameters associated with the tissue, bath media, and impedance were identified. Variables were defined experimentally and/or from literature sources to best mimic the clinical cardiac ablation setting. The model was validated by performing three independent replicates of ex vivo myocardial ablation and a direct comparison of lesion outcomes of the ex vivo swine myocardial and in vivo canine thigh preparation (TP) models. Replicate experiments on the ex vivo model demonstrated low variance in ablation depth (6.5 ± 0.6, 6.3 ± 0.6, 6.2 ± 0.4 mm) and width (10.4 ± 1.1, 9.7 ± 1.0, 9.9 ± 0.9 mm) and no significant differences between replicates. In a direct comparison of the two models, the ex vivo model demonstrated ablation depths similar to the canine TP model at 35 W (6.9 ± 1.0, and 7.0 ± 0.9 mm) and 50 W (8.0 ± 0.7, and 8.4 ± 0.7 mm), as well as similar power to depth ratios (15% and 19% for the ex vivo cardiac and in vivo TP models, respectively). CONCLUSION: The ex vivo model exhibited strong lesion reproducibility and power-to-depth ratios comparable to the in vivo TP model. The optimized ex vivo model minimizes animal usage with increased throughput, lesion characteristics similar to the in vivo TP model, and ability to discriminate minor variations between different catheter designs.
No takes yet. Share an insight, caveat, or question.
Lacko et al. (2022) studied Cardiac ablation. Ex vivo swine myocardial ablation model vs. In vivo canine thigh preparation model was evaluated on Ablation depth and width. An optimized ex vivo swine myocardial ablation model demonstrated ablation depths similar to an in vivo canine thigh preparation model at 35 W and 50 W.
Synapse has enriched 10 related papers. Consider them for comparative context: