Key result
Normalized intrinsic left ventricular electric delay (QLV/QRSd) was associated with acute hemodynamic response (P<0.05) but cannot be used to select the optimal electrode in individual patients.
Why the study?
Does the intrinsic left ventricular delay (QLV) optimize the pacing configuration for acute hemodynamic response in heart failure patients receiving CRT with a quadripolar LV lead?
Observational (n=48)
Does the intrinsic left ventricular delay (QLV) optimize the pacing configuration for acute hemodynamic response in heart failure patients receiving CRT with a quadripolar LV lead?
Although QLV normalized for QRS duration is associated with acute hemodynamic response at the group level, it cannot be used to select the optimal electrode in individual patients receiving CRT with a quadripolar lead.
QLV/QRSd should not guide electrode selection in individual quadripolar CRT patients; leaves open whether it aids personalized hemodynamic optimization.
Background: Previous studies indicated the importance of the intrinsic left ventricular (LV) electric delay (QLV) for optimal benefit to cardiac resynchronization therapy. We investigated the use of QLV for achieving optimal acute hemodynamic response to cardiac resynchronization therapy with a quadripolar LV lead. Methods and Results: Forty-eight heart failure patients with a left bundle branch block were prospectively enrolled (31 men; age, 66±10 years; LV ejection fraction, 28±8%; QRS duration, 176±14 ms). Immediately after cardiac resynchronization therapy implantation, invasive LV pressure–volume loops were recorded during biventricular pacing with each separate electrode at 4 atrioventricular delays. Acute cardiac resynchronization therapy response, measured as change in stroke work (Δ%SW) compared with intrinsic conduction, was related to intrinsic interval between Q on the ECG and LV sensing delay (QLV), normalized for QRS duration (QLV/QRSd), and electrode position. QLV/QRSd was 84±9% and variation between the 4 electrodes 9±5%. Δ%SW was 89±64% and varied by 39±36% between the electrodes. In univariate analysis, an anterolateral or lateral electrode position and a high QLV/QRSd had a significant association with a large Δ%SW (all P <0.01). In a combined model, only QLV/QRSd remained significantly associated with Δ%SW ( P <0.05). However, a direct relation between QLV/QRSd and Δ%SW was only seen in 24 patients, whereas 24 patients showed an inverse relation. Conclusions: The large variation in acute hemodynamic response indicates that the choice of the stimulated electrode on a quadripolar lead is important. Although QLV/QRSd was associated with acute hemodynamic response at group level, it cannot be used to select the optimal electrode in the individual patient.
No takes yet. Share an insight, caveat, or question.
Everdingen et al. (2018) conducted an observational in Heart failure with left bundle branch block (n=48). Biventricular pacing with quadripolar LV lead vs. Intrinsic conduction was evaluated on Acute hemodynamic response measured as change in stroke work (Δ%SW) (p=<0.05). Normalized intrinsic left ventricular electric delay (QLV/QRSd) was associated with acute hemodynamic response (P<0.05) but cannot be used to select the optimal electrode in individual patients.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: