Key result
Quadripolar left ventricular pacing configurations increased cardiac index by for unipolar-simulated configurations, and yielded a 90% CRT responder rate.
Why the study?
Does quadripolar pacing configuration improve hemodynamic response in patients implanted with a CRT device compared to unipolar-simulated and bipolar configurations?
Does quadripolar pacing configuration improve hemodynamic response in patients implanted with a CRT device compared to unipolar-simulated and bipolar configurations?
Quadripolar LV pacing configurations significantly improve hemodynamic response and CRT responder rates compared to unipolar and bipolar configurations.
Answers are grounded in this paper and cite their sources.
Idea: Quadripolar LV pacing may optimize hemodynamics.
BACKGROUND: The objective of the present study was to evaluate the usefulness of a left ventricle (LV) quadripolar lead in improving the hemodynamic response to cardiac resynchronization therapy (CRT). METHODS AND RESULTS: We included 27 consecutive patients implanted with a CRT device with an LV quadripolar lead. Hemodynamic parameters were evaluated at 3-month follow-up by using impedance cardiography. We assessed the highest cardiac output and the highest stroke volume (SV) obtained after atrioventricular and interventricular optimization with pacing from each of the four electrodes of the LV lead. Each patient was evaluated according to three different pacing configurations: unipolar-simulated, bipolar, and quadripolar. Biventricular pacing improved hemodynamics in comparison to the nonpaced measurements: cardiac index (CI): 2.69 L/min/m(2) versus 2.17 L/min/m(2) (P = 0.001). The hemodynamic response was highest in the quadripolar in comparison to unipolar-simulated and bipolar configurations, with an increase of 29%, 23%, and 27%, respectively, in relation to the reference CI and with an increase of 22%, 11%, and 18%, respectively, in relation to the reference indexed SV (P < 0.05, for the comparison between unipolar-simulated and quadripolar configurations). Twelve patients (44%) showed the best hemodynamic response by pacing from any of the two distal electrodes and 15 patients (56%) by pacing from the two proximal electrodes. Finally, CRT responder rates were higher in quadripolar versus bipolar and unipolar-simulated configurations: 90%, 85%, and 75%, respectively. CONCLUSIONS: The quadripolar LV lead was associated with a better hemodynamic response and higher CRT responder rates when compared with unipolar-simulated and bipolar LV leads.
No takes yet. Share an insight, caveat, or question.
Osca et al. (2014) studied Patients implanted with a CRT device with an LV quadripolar lead (n=27). Quadripolar pacing configuration vs. Unipolar-simulated and bipolar pacing configurations was evaluated on Increase in cardiac index in relation to reference (p=<0.05). Quadripolar left ventricular pacing configurations increased cardiac index by for unipolar-simulated configurations, and yielded a 90% CRT responder rate.