Key result
Individually optimizing the pacing configuration of a quadripolar left ventricular lead produced an absolute 10.0% additional increase in %DLV+dP/dtmax compared to the worst pacing vector.
Why the study?
Does individualized vector selection of a quadripolar left ventricular pacing lead improve acute hemodynamic response in patients receiving cardiac resynchronization therapy?
RCT (n=16)
Treatment concealment by uninvolved staff
Computer-generated randomization list
No
Does individualized vector selection of a quadripolar left ventricular pacing lead improve acute hemodynamic response in patients receiving cardiac resynchronization therapy?
Individualized vector selection using a quadripolar left ventricular lead significantly improves acute hemodynamic response compared to default or worst pacing configurations in patients undergoing cardiac resynchronization therapy.
Answers are grounded in this paper and cite their sources.
Individualized quadripolar LV lead optimization improves acute CRT hemodynamics; extends vector selection evidence and supports outcome trials.
BACKGROUND: A suboptimal left ventricular (LV) pacing site may account for non-responsiveness of patients to cardiac resynchronization therapy (CRT). The vector selection of a novel quadripolar LV pacing lead, which was mainly developed to overcome technical issues with stimulation thresholds and phrenic nerve capture, may affect hemodynamic response, and was therefore assessed in this study. (German Clinical Trials Register DRKS00000573). METHODS AND RESULTS: Hemodynamic effects of a total of 145 LVPCs (9.1 per patient) of CRT devices with a quadripolar LV lead (Quartet™, St. Jude Medical) were assessed in 16/20 consecutive patients by invasive measurement of LV+dP/dtmax at an invasively optimized AV-interval in random order. Optimal (worst) LVPCs per patient were identified as those with maximal (minimal) %change in LV+dP/dtmax (%ΔLV+dP/dtmax) as compared to a preceding baseline. LV+dP/dtmax significantly increased in all 145 LVPCs (p<0.0001 compared to baseline) with significant intraindividual differences between LVPCs (p<0.0001). Overall, CRT acutely augmented %ΔLV+dP/dtmax by 31.3% (95% CI 24%-39%) in the optimal, by 21.3% (95% CI: 15%-27%) in the worst and by 28.2% (95% CI: 21%-36%) in a default distal LVPC. This resulted in an absolute additional acute increase in %ΔLV+dP/dtmax of 10.0% (95% CI: 7%-13%) of the optimal when compared to the worst (p<0.0001), and of 3.1% (95% CI: 1%-5%) of the optimal when compared to the default distal LVPC (p<0.001). Optimal LVPCs were not programmable with a standard bipolar lead in 44% (7/16) of patients. CONCLUSION: The pacing configuration of a quadripolar LV lead determinates acute hemodynamic response. Pacing in the individually optimized configuration gives rise to an additional absolute 10% increase in %ΔLV+dP/dtmax when comparing optimal and worst vectors.
No takes yet. Share an insight, caveat, or question.
Asbach et al. (2013) conducted an RCT in Heart Failure (n=16). Individually optimized left ventricular pacing configuration (LVPC) vs. Worst LVPC was evaluated on Acute change in LV +dP/dtmax (%DLV+dP/dtmax) from baseline (MD 10.0%, 95% CI 7%-13%, p=<0.0001). Individually optimizing the pacing configuration of a quadripolar left ventricular lead produced an absolute 10.0% additional increase in %DLV+dP/dtmax compared to the worst pacing vector.
Synapse has enriched 10 related papers. Consider them for comparative context: