Key result
A quadripolar left ventricular lead reduced intraoperative pacing thresholds compared to standard bipolar leads, enabling successful biventricular pacing in 92%.
Why the study?
Does a quadripolar LV lead improve successful biventricular pacing rates and lower pacing thresholds compared to standard bipolar leads in CRT patients with intraoperative PNS or high pacing thresholds?
Observational (n=26)
Does a quadripolar LV lead improve successful biventricular pacing rates and lower pacing thresholds compared to standard bipolar leads in CRT patients with intraoperative PNS or high pacing thresholds?
In patients undergoing CRT who experience phrenic nerve stimulation or high pacing thresholds with conventional bipolar leads, a quadripolar LV lead significantly increases the rate of successful biventricular stimulation and lowers pacing thresholds.
Answers are grounded in this paper and cite their sources.
May allow CRT implantation when bipolar leads cause PNS or high thresholds; leaves open need for outcome trials.
BACKGROUND: Phrenic nerve stimulation (PNS) and high pacing thresholds (HPT) hinder biventricular stimulation in patients (pts) undergoing cardiac resynchronization therapy (CRT). A new quadripolar left ventricular (LV) lead (Quartet 1458Q, St. Jude Medical) with increased number of pacing configuration, might overcome this problem. METHODS: All consecutive pts in whom a standard bipolar lead intraoperatively resulted in PNS and/or HPT (≥4.00V/1mV), received, during the same implant, a quadripolar LV lead. Aim of the study was to evaluate acute and short term outcome. RESULTS: 26 pts [24 (92%) male, mean age 74±6 years)] with PNS (22 pts; 85%) and HPT (4 pts; 15%) were included. Permanent right ventricular pacing was the reason for broad QRS complex in 4 (15%) pts, whereas all other pts had a left bundle branch block. Severely symptomatic (NYHA Class ≥3) heart failure with reduced ejection fraction (EF 31±9%) was mostly caused by ischemic heart disease (14 pts; 54%). Idiopathic dilated cardiomyopathy and valvular heart disease were diagnosed in 6 (23%) pts each. In most (24/26, 92%) pts the use of the Quartet lead led to successful biventricular pacing due to a significant reduction in intraoperative pacing threshold (5.2V/1.0ms vs. 1.4V/0.8ms; p=0.03), which was maintained (1.2V/0.7ms) at follow-up. PNS never represented reason for failed LV pacing, neither acutely nor during follow-up. CONCLUSION: Excessively HPT and/or PNS are frequently encountered when conventional bipolar leads are used for CRT. A new quadripolar LV lead increases the rate of successful biventricular stimulation. Lower pacing threshold and freedom from PNS are maintained at follow-up.
No takes yet. Share an insight, caveat, or question.
Ohlow et al. (2013) conducted an observational in Heart failure undergoing cardiac resynchronization therapy (n=26). Quadripolar left ventricular lead (Quartet 1458Q) vs. Standard bipolar lead was evaluated on Intraoperative pacing threshold (V) (p=0.03). A quadripolar left ventricular lead reduced intraoperative pacing thresholds compared to standard bipolar leads, enabling successful biventricular pacing in 92%.
Synapse has enriched 10 related papers. Consider them for comparative context: