Key result
The St Jude Quartet quadripolar LV lead achieved successful cardiac resynchronization therapy in 4 out of 4 patients with previous failed attempts, overcoming phrenic nerve stimulation.
Why the study?
Does a quadripolar left ventricular lead enable successful CRT implantation in patients with previous failed attempts?
Case Report (n=4)
Does a quadripolar left ventricular lead enable successful CRT implantation in patients with previous failed attempts?
A quadripolar LV lead can successfully salvage CRT implantation in patients who previously failed due to phrenic nerve stimulation or high capture thresholds.
Answers are grounded in this paper and cite their sources.
Quadripolar LV leads may enable CRT after prior failures; leaves open durability and generalizability beyond small series.
AIMS: Problems with implanting a left ventricular (LV) lead during cardiac resynchronization therapy (CRT) procedures are not uncommon and may occur for a variety of reasons including phrenic nerve stimulation (PNS) and high capture thresholds. We aimed to perform successful CRT in patients with previous LV lead problems using the multiple pacing configurations available with the St Jude Quartet model 1458Q quadripolar LV lead to overcome PNS or high capture thresholds. METHODS AND RESULTS: Four patients with previous failed attempts at LV lead implantation underwent a further attempt at CRT using a Quartet lead. In all four cases, successful CRT was achieved using a Quartet lead placed in a branch of the coronary sinus. Problems with PNS or high capture thresholds were seen in all four patients but were successfully overcome. Satisfactory lead parameters were seen at implant, pre-discharge, and at short-term follow-up (8.5±5 weeks). CONCLUSION: The Quartet lead allows 10 different pacing vectors to be used and may overcome common pacing problems because of the multiple pacing configurations available. Problems with either PNS or unsatisfactory pacing parameters experienced during CRT may be resolved simply by changing the pacing configuration using this quadripolar lead system.
No takes yet. Share an insight, caveat, or question.
Shetty et al. (2011) conducted a case report in previous failed attempts at cardiac resynchronization therapy (n=4). St Jude Quartet model 1458Q quadripolar LV lead was evaluated on successful CRT. The St Jude Quartet quadripolar LV lead achieved successful cardiac resynchronization therapy in 4 out of 4 patients with previous failed attempts, overcoming phrenic nerve stimulation.