Key result
The Quartet LV lead demonstrated stable electrical performance, a 95.5% implant success rate, a 3.5% dislodgement rate, and enabled noninvasive elimination of phrenic nerve stimulation in all cases.
Why the study?
Does the Quartet quadripolar LV lead provide stable electrical performance and allow noninvasive resolution of phrenic nerve stimulation in patients requiring CRT-D?
Does the Quartet quadripolar LV lead provide stable electrical performance and allow noninvasive resolution of phrenic nerve stimulation in patients requiring CRT-D?
Answers are grounded in this paper and cite their sources.
The quadripolar Quartet LV lead demonstrates stable electrical performance and allows for noninvasive resolution of phrenic nerve stimulation through multiple pacing vectors.
INTRODUCTION: The Quartet(®) left ventricular (LV) lead is the first with 4 pacing electrodes (tip and 3 rings) that enables pacing from 10 different pacing vectors. Postoperative performance of this lead was evaluated in a prospective, nonrandomized, multicenter IDE study. METHODS: Patients with standard indications for CRT-D were enrolled. Electrical performance and presence of phrenic nerve stimulation (PNS) were assessed during pacing from each of 10 vectors at predischarge (within 7 days), 1 month, and 3 months postimplant. RESULTS: The Quartet LV lead was implanted successfully in 170 patients (95.5% implant success rate, 68 ± 11 years, 68.5% male, LVEF: 25 ± 7%, NYHA class III: 98.3% and class IV: 1.7%). Mean follow-up was 4.7 ± 1.9 months. Capture threshold and impedance for each of the 10 LV lead pacing vectors remained stable during follow-up. LV lead dislodgement occurred in 6 (3.5%) patients and PNS was observed in 23 (13.5%) patients. PNS was resolved noninvasively in all 23 (100%) patients, either by reprogramming to pace from the additional LV lead pacing vectors alone (13 pts, 56.5%), reprogramming to pace from the additional LV lead pacing vectors and reprogramming pacing output (4 pts, 17.4%), or by reprogramming pacing output alone (6 pts, 26.1%). CONCLUSIONS: The Quartet LV lead electrical performance was stable and was associated with a high implant success and low dislodgement rate during 3-month follow-up. In all patients with PNS, the 10 pacing vectors combined with reduced output programming enabled the elimination of PNS noninvasively.
No takes yet. Share an insight, caveat, or question.
Tomassoni et al. (2013) studied Standard indications for CRT-D (n=170). Quartet left ventricular (LV) lead was evaluated on Electrical performance and presence of phrenic nerve stimulation (PNS). The Quartet LV lead demonstrated stable electrical performance, a 95.5% implant success rate, a 3.5% dislodgement rate, and enabled noninvasive elimination of phrenic nerve stimulation in all cases.