Key result
Inverse thermodilution using conventional pulmonary artery catheters correlated well with ultrasound Doppler for renal flow but not for cerebral or femoral flows.
Why the study?
Does inverse thermodilution using conventional pulmonary artery catheters accurately measure regional venous blood flows compared to ultrasound Doppler in an animal model?
Does inverse thermodilution using conventional pulmonary artery catheters accurately measure regional venous blood flows compared to ultrasound Doppler in an animal model?
Inverse thermodilution using conventional PACs is not a reliable method for measuring most regional venous blood flows in vivo due to poor correlation and significant technical limitations.
Answers are grounded in this paper and cite their sources.
Does not support clinical use of inverse thermodilution for most regional flows; leaves open targeted validation in humans.
Assessment of regional blood flow changes is difficult in the clinical setting. We tested whether conventional pulmonary artery catheters (PACs) can be used to measure regional venous blood flows by inverse thermodilution (ITD). Inverse thermodilution was tested in vitro and in vivo using perivascular ultrasound Doppler (USD) flow probes as a reference. In anesthetized pigs, PACs were inserted in jugular, hepatic, renal, and femoral veins, and their measurements were compared with simultaneous USD flow measurements from carotid, hepatic, renal, and femoral arteries and from portal vein. Fluid boluses were injected through the PAC's distal port, and temperature changes were recorded from the proximally located thermistor. Injectates of 2 and 5 mL at 22 degrees C and 4 degrees C were used. Flows were altered by using a roller pump (in vitro), and infusion of dobutamine and induction of cardiac tamponade, respectively. In vitro: At blood flows between 400 mL . min-1 and 700 mL . min-1 (n = 50), ITD and USD correlated well (r = 0.86, P < 0.0001), with bias and limits of agreement of 3 +/- 101 mL . min-1. In vivo: 514 pairs of measurements had to be excluded from analysis for technical reasons, and 976 were analyzed. Best correlations were r = 0.87 (P < 0.0001) for renal flow and r = 0.46 (P < 0.0001) for hepatic flow. No significant correlation was found for cerebral and femoral flows. Inverse thermodilution using conventional PAC compared moderately well with USD for renal but not for other flows despite good in vitro correlation in various conditions. In addition, this method has significant technical limitations.
No takes yet. Share an insight, caveat, or question.
Ganter et al. (2009) studied Regional blood flow assessment. Inverse thermodilution using conventional pulmonary artery catheters vs. Perivascular ultrasound Doppler flow probes was evaluated on Correlation between inverse thermodilution and ultrasound Doppler measurements (r = 0.87 (renal flow), p=<0.0001). Inverse thermodilution using conventional pulmonary artery catheters correlated well with ultrasound Doppler for renal flow but not for cerebral or femoral flows.