Key result
Ventricular decremental stimulation revealed the minimum cycle length maintaining 1:1 VA conduction was longer than the VT cycle (496 vs 320 msec; P<0.01), with no 1:1 VA conduction during VT.
Observational (n=66)
1:1 ventriculoatrial conduction during ventricular tachycardia is rare in ICD recipients, suggesting it is unlikely to complicate dual-chamber tachyarrhythmia detection algorithms.
Incorporation of atrial electrograms in the tachycardia detection algorithm may improve tachyarrhythmia discrimination by ICDs but retrograde ventriculoatrial (VA) conduction over the AV node during ventricular tachyarrhythmia may be problematic. The present study analyzed VA conduction characteristics in 66 ICD patients who had evaluation of the VA conduction system by electrophysiological studies before implant. VA conduction was demonstrated in patients during ventricular decremental stimulation. Forty patients had inducible sustained monomorphic VT. The minimum cycle length maintaining 1:1 VA conduction during ventricular stimulation was longer than the cycle of VT in every patient (496 +/- 100 msec vs 320 +/- 81 msec; P < 0.01). Occasional VA conduction during VT was observed in five patients and one patient had 2:1 VA conduction during induced VT. No patient had 1:1 VA conduction during VT. We conclude that brisk VA conduction is uncommon and 1:1 VA conduction during VT is rare ICD recipients. VA conduction is unlikely to complicate the incorporation of atrial electrograms into tachyarrhythmia detection algorithms.
No takes yet. Share an insight, caveat, or question.
LI et al. (1994) conducted an observational in Implantable Cardioverter Defibrillator (ICD) patients (n=66). Ventricular decremental stimulation vs. Induced sustained monomorphic VT was evaluated on Minimum cycle length maintaining 1:1 VA conduction (p=<0.01). Ventricular decremental stimulation revealed the minimum cycle length maintaining 1:1 VA conduction was longer than the VT cycle (496 vs 320 msec; P<0.01), with no 1:1 VA conduction during VT.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: