Key result
Placing clip-on and closed-back headphones within 3 mm temporarily suspended tachyarrhythmia detection and therapy in 1 of 3 ICDs tested, an effect not observed at distances of at least 2 cm.
Why the study?
Does exposure to static magnetic fields from earphones and headphones interfere with the function of pacemakers and ICDs?
Does exposure to static magnetic fields from earphones and headphones interfere with the function of pacemakers and ICDs?
Patients with ICDs should keep earphones and headphones at least 2 cm away from their devices to prevent temporary suspension of tachyarrhythmia detection and therapy.
Answers are grounded in this paper and cite their sources.
Close headphone proximity may suspend ICD detection in select animal models; leaves open human relevance at typical distances.
In this study, we evaluated the effects of static magnetic fields of earphones and headphones on pacemakers and implantable cardioverter defibrillators(ICDs). Five pacemakers and three ICDs were subjected to in-vitro test with three headphones which were an in-ear earphone, clip-on headphone, and closed-back headphone. Each implantable device was placed in close proximity(within 3 mm) to the ear-pad of each of the earphone and headphones for 3 min. As a result, no effects were observed on the pacemakers for the earphone and headphones during the test, but an effect was observed on one ICD for the clip-on and closed-back headphone during the test. When the ICD was placed in close proximity to the headphones, the ICD temporarily suspended functions of tachyarrhythmia detection and therapy. The effect was not observed in this study when the headphones were at least 2 cm from the ICD. Based on these findings, patients with ICDs should be advised to keep earphones and headphones at least 2 cm apart from their ICDs.
No takes yet. Share an insight, caveat, or question.
Chung et al. (2015) studied Pacemakers and implantable cardioverter defibrillators (n=8). Earphones and headphones (in-ear, clip-on, closed-back) was evaluated on Effects of static magnetic fields on device function. Placing clip-on and closed-back headphones within 3 mm temporarily suspended tachyarrhythmia detection and therapy in 1 of 3 ICDs tested, an effect not observed at distances of at least 2 cm.