Key result
In vitro, 26-mm Evolut R valves in small aortas yield ~42% higher peak pressure gradients.
Why the study?
The study aimed to evaluate how interaction between the Medtronic Evolut R transcatheter aortic valve and the ascending aorta affects valve performance and pressure recovery across different ascending aorta diameters.
Does ascending aorta size affect the pressure gradient and pressure recovery of a self-expandable transcatheter aortic valve in an in vitro model?
Does ascending aorta size affect the pressure gradient and pressure recovery of a self-expandable transcatheter aortic valve in an in vitro model?
Absolute Event Rate: 11.5% vs 8.1%
p-value: p=<.01
In an in vitro model, ascending aorta size significantly impacts the pressure gradient and recovery of a self-expandable transcatheter aortic valve, suggesting AA size should be considered during device selection.
Small ascending aorta size may raise Evolut R gradients; hypothesis-generating for device selection and requires clinical validation.
ObjectiveIn this study we aimed to understand the role of interaction of the Medtronic Evolut R transcatheter aortic valve with the ascending aorta (AA) by evaluating the performance of the valve and the pressure recovery in different AA diameters with the same aortic annulus size.MethodsA 26-mm Medtronic Evolut R valve was tested using a left heart simulator in aortic root models of different AA diameter (D): small (D = 23 mm), medium (D = 28 mm), and large (D = 34 mm) under physiological conditions. Measurements of pressure from upstream to downstream of the valve were performed using a catheter at small intervals to comprehensively assess pressure gradient and pressure recovery.ResultsIn the small AA, the measured peak and mean pressure gradient at vena contracta were 11.5 ± 0.5 mm Hg and 7.8 ± 0.4 mm Hg, respectively, which was higher (P < .01) compared with the medium (8.1 ± 0.4 mm Hg and 5.2 ± 0.4 mm Hg) and large AAs (7.4 ± 1.0 mm Hg and 5.4 ± 0.6 mm Hg). The net pressure gradient was lower for the case with the medium AA (4.1 ± 1.2 mm Hg) compared with the small AA (4.7 ± 0.8 mm Hg) and large AA (6.1 ± 1.4 mm Hg; P < .01).ConclusionsWe have shown that small and large AAs can increase net pressure gradient, because of the direct interaction of the Medtronic Evolut R stent with the AA (in small AA) and introducing higher level of turbulence (in large AA). AA size might need to be considered in the selection of an appropriate device for transcatheter aortic valve replacement.
No takes yet. Share an insight, caveat, or question.
Samaee et al. (2022) studied Aortic stenosis (n=156). Medtronic Evolut R transcatheter aortic valve vs. Different ascending aorta sizes (small, medium, large) was evaluated on Peak pressure gradient at vena contracta (p=<.01). In an in vitro model, the 26-mm Medtronic Evolut R valve deployed in a small ascending aorta had a significantly higher peak pressure gradient (11.5 mm Hg) compared to medium (8.1 mm Hg) and large (7.4 mm Hg) aortas.