Key result
Aktiia Bracelet meets ISO validation criteria versus auscultation over one month with ~0.5 mmHg mean difference.
Why the study?
The study was conducted to assess the accuracy and precision of the Aktiia Bracelet, a noninvasive optical wrist blood pressure monitor, over 1 month.
Does the Aktiia Bracelet accurately and precisely measure blood pressure compared to double-blinded auscultation in adults over 1 month?
Observational (n=86)
Double-blind
No
Does the Aktiia Bracelet accurately and precisely measure blood pressure compared to double-blinded auscultation in adults over 1 month?
Effect estimate: SBP 0.46 ± 7.75 mmHg, DBP 0.39 ± 6.86 mmHg
The Aktiia Bracelet, a cuffless optical wrist blood pressure monitor, demonstrated acceptable accuracy and precision over 1 month compared to standard auscultation, satisfying ISO81060-2 criteria in the sitting position.
Supports further cuffless BP device evaluation; leaves open generalizability and outcome impact in routine practice.
OBJECTIVE: The objective of this study (NCT04027777) was to assess the accuracy and precision of the Aktiia Bracelet, a CE-marked noninvasive optical blood pressure (BP) monitor worn at the wrist, over a period of 1 month. METHODS: In this study, participants aged between 21 and 65 years were recruited. The clinical investigation extended the ISO81060-2:2013 standard to the specificities of cuffless devices. Each BP assessment consisted of the simultaneous recording of optical signals with Aktiia Bracelet and double-blinded auscultation by two trained observers in the standard sitting position. The algorithms of Aktiia Bracelet further processed the recorded optical signals to perform a signal quality check and to calculate uncalibrated estimates of systolic BP (SBP) and diastolic BP (DBP). These estimates were transformed into mmHg using a subject-dependent calibration parameter, which was calculated using the first two available reference measurements per subject. RESULTS: Eighty-six participants were included in the analysis. The mean and SD of the differences between Aktiia Bracelet estimates and the reference (ISO81060-2 criterion 1) were 0.46 ± 7.75 mmHg for SBP and 0.39 ± 6.86 mmHg for DBP. The SD of the averaged paired difference per subject (ISO81060-2 criterion 2) were 3.9 mmHg for SBP and 3.6 mmHg for DBP. CONCLUSION: After initialization and during 1 month, the overall accuracy of Aktiia Bracelet satisfied validation criteria 1 and 2 of ISO81060-2 in the sitting position. The Aktiia Bracelet can be recommended for BP measurement in the adult population.
No takes yet. Share an insight, caveat, or question.
Vybornova et al. (2021) conducted an observational in Blood pressure monitoring (n=86). Aktiia Bracelet vs. Double-blinded auscultation was evaluated on Mean difference between Aktiia Bracelet and reference (ISO81060-2 criterion 1) (SBP 0.46 ± 7.75 mmHg, DBP 0.39 ± 6.86 mmHg). The Aktiia Bracelet satisfied ISO81060-2 validation criteria over 1 month, with mean differences compared to auscultation of 0.46 ± 7.75 mmHg for SBP and 0.39 ± 6.86 mmHg for DBP.