Key result
12-lead ECG and automated screening differentiate true S-ICD lead fractures from pseudo-fractures during alternate vector flatlines.
Why the study?
Does a flatline on the alternate vector (AVFL) indicate potential electrode body fracture in patients with an S-ICD?
Case Report (n=2)
No
Does a flatline on the alternate vector (AVFL) indicate potential electrode body fracture in patients with an S-ICD?
Answers are grounded in this paper and cite their sources.
A flatline on the alternate vector during S-ICD interrogation may serve as an early and potentially sole indicator of electrode body fracture.
In February 2021 the United States Food and Drug Administration (FDA) issued a Class I recall of Boston Scientific’s Emblem subcutaneous implantable cardiac defibrillator (S-ICD) model 3501 electrode (Boston Scientific, Marlborough, MA) following 27 cases of lead fracture.1 These fractures may be detected by non-physiological artefacts or high impedance alerts according to device programming.2 A flatline on the alternate vector (AVFL) on in-office check has further been highlighted by Boston Scientific as an important early (and possible sole) indicator of potential electrode body fracture.
No takes yet. Share an insight, caveat, or question.
O’Neill et al. (2021) conducted a case report in Ischemic cardiomyopathy with subcutaneous implantable cardiac defibrillator (S-ICD) (n=2). Subcutaneous implantable cardiac defibrillator (S-ICD) was evaluated. An alternate vector flatline on S-ICD interrogation may indicate a true lead fracture or a pseudo-fracture, which can be differentiated using surface 12-lead ECG and the automated screening tool.