The proposed wearable cuff-less single-arm ECG and PPG monitoring system estimated systolic blood pressure with a mean error of 1.63 ± 4.44 mmHg and a mean absolute error of 3.68 mmHg.
Does a wearable single-arm ECG and PPG-based system accurately estimate systolic blood pressure and heart rate compared to reference methods in volunteers?
A novel wearable single-arm ECG and PPG system using a combined PTT and HR model demonstrated high accuracy for continuous cuff-less blood pressure and heart rate monitoring in a proof-of-concept study.
BACKGROUND: Long-term continuous systolic blood pressure (SBP) and heart rate (HR) monitors are of tremendous value to medical (cardiovascular, circulatory and cerebrovascular management), wellness (emotional and stress tracking) and fitness (performance monitoring) applications, but face several major impediments, such as poor wearability, lack of widely accepted robust SBP models and insufficient proofing of the generalization ability of calibrated models. METHODS: This paper proposes a wearable cuff-less electrocardiography (ECG) and photoplethysmogram (PPG)-based SBP and HR monitoring system and many efforts are made focusing on above challenges. Firstly, both ECG/PPG sensors are integrated into a single-arm band to provide a super wearability. A highly convenient but challenging single-lead configuration is proposed for weak single-arm-ECG acquisition, instead of placing the electrodes on the chest, or two wrists. Secondly, to identify heartbeats and estimate HR from the motion artifacts-sensitive weak arm-ECG, a machine learning-enabled framework is applied. Then ECG-PPG heartbeat pairs are determined for pulse transit time (PTT) measurement. Thirdly, a PTT&HR-SBP model is applied for SBP estimation, which is also compared with many PTT-SBP models to demonstrate the necessity to introduce HR information in model establishment. Fourthly, the fitted SBP models are further evaluated on the unseen data to illustrate the generalization ability. A customized hardware prototype was established and a dataset collected from ten volunteers was acquired to evaluate the proof-of-concept system. RESULTS: The semi-customized prototype successfully acquired from the left upper arm the PPG signal, and the weak ECG signal, the amplitude of which is only around 10% of that of the chest-ECG. The HR estimation has a mean absolute error (MAE) and a root mean square error (RMSE) of only 0.21 and 1.20 beats per min, respectively. Through the comparative analysis, the PTT&HR-SBP models significantly outperform the PTT-SBP models. The testing performance is 1.63 ± 4.44, 3.68, 4.71 mmHg in terms of mean error ± standard deviation, MAE and RMSE, respectively, indicating a good generalization ability on the unseen fresh data. CONCLUSIONS: The proposed proof-of-concept system is highly wearable, and its robustness is thoroughly evaluated on different modeling strategies and also the unseen data, which are expected to contribute to long-term pervasive hypertension, heart health and fitness management.
Zhang et al. (Mon,) 在健康志愿者 (n=10) 中进行了一项其他研究。在收缩压估计平均误差 (PTT&HR-SBP model 10) 方面评估了可穿戴无袖带单臂 ECG 和 PPG 监测系统对比基于袖带的 SBP 和胸部 ECG。所提出的可穿戴无袖带单臂 ECG 和 PPG 监测系统估计收缩压的平均误差为 1.63 ± 4.44 mmHg,平均绝对误差为 3.68 mmHg。
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: