Key result
Pre-dilation balloon under-expansion on intravascular imaging strongly predicts target lesion stiffness with >0.89 correlation.
Why the study?
Stent-induced mechanical stimuli cause post-treatment pathophysiological responses, but the lack of target lesion material characterization limits personalized treatment.
Does OCT-integrated angioplasty balloon under-expansion predict hyperelastic behavior and local stiffness of coronary artery lesions in ex vivo human donor hearts?
Observational (n=6)
No
Does OCT-integrated angioplasty balloon under-expansion predict hyperelastic behavior and local stiffness of coronary artery lesions in ex vivo human donor hearts?
Effect estimate: r > 0.89
p-value: p=<0.001
Intravascular OCT imaging of balloon under-expansion during pre-dilation can predict target lesion stiffness, offering a potential method to personalize stent deployment.
May support OCT-guided lesion assessment; hypothesis-generating in ex vivo models, leaves open in vivo validation.
Introduction: Stent-induced mechanical stimuli cause pathophysiological responses in the coronary artery post-treatment. These stimuli can be minimized through choice of stent, size, and deployment strategy. However, the lack of target lesion material characterization is a barrier to further personalizing treatment. A novel ex-vivo angioplasty-based intravascular imaging technique using optical coherence tomography (OCT) was developed to characterize local stiffness of the target lesion. Methods: After proper institutional oversight, atherosclerotic coronary arteries ( n = 9) were dissected from human donor hearts for ex vivo material characterization <48 h post-mortem. Morphology was imaged at the diastolic blood pressure using common intravascular OCT protocols and at subsequent pressures using a specially fabricated perfusion balloon that accommodates the OCT imaging wire. Balloon under-expansion was quantified relative to the nominal balloon size at 8 ATM. Correlation to a constitutive hyperelastic model was empirically investigated ( n = 13 plaques) using biaxial extension results fit to a mixed Neo-Hookean and Exponential constitutive model. Results and discussion: The average circumferential Cauchy stress was 66.5, 130.2, and 300.4 kPa for regions with <15, 15–30, and >30% balloon under-expansion at a 1.15 stretch ratio. Similarly, the average longitudinal Cauchy stress was 68.1, 172.6, and 412.7 kPa, respectively. Consequently, strong correlation coefficients >0.89 were observed between balloon under-expansion and stress-like constitutive parameters. These parameters allowed for visualization of stiffness and material heterogeneity for a range of atherosclerotic plaques. Balloon under-expansion is a strong predictor of target lesion stiffness. These findings are promising as stent deployment could now be further personalized via target lesion material characterization obtained pre-operatively.
No takes yet. Share an insight, caveat, or question.
Ghorbannia et al. (2023) conducted an observational in Coronary artery disease (n=6). OCT-integrated angioplasty balloon was evaluated on Correlation between balloon under-expansion and stress-like constitutive parameters (r > 0.89, p=<0.001). Intravascular imaging of angioplasty balloon under-expansion during pre-dilation strongly predicted target lesion stiffness, with correlation coefficients >0.89.