Why the study?
The SMART Pass filter reduces inappropriate shocks from subcutaneous implantable cardioverter-defibrillators but is deactivated when R-wave sensing is diminished.
Massive intra-abdominal hemorrhage can alter the S-ICD sensing vector, leading to reduced R-wave sensing, automatic SMART Pass filter deactivation, and subsequent inappropriate shocks due to T-wave oversensing.
Answers are grounded in this paper and cite their sources.
Extensive hemorrhage may deactivate SMART Pass and provoke inappropriate S-ICD shocks; hypothesis-generating case report requiring prospective validation.
The SMART Pass filter (Boston Scientific) aims to reduce inappropriate shocks (IASs) from subcutaneous implantable cardioverter-defibrillators by filtering out low-frequency signals such as T waves. However, this filter is deactivated in the presence of diminished R-wave sensing. We describe a case of IAS in the setting of extensive intra-abdominal hemorrhage.
No takes yet. Share an insight, caveat, or question.
Malaty et al. (2023) studied this question.
Synapse has enriched 10 related papers. Consider them for comparative context: