Why the study?
Does reducing potassium in the dialysis bath resolve hyperkalemia-induced T wave oversensing and inappropriate ICD shocks in a patient with a biventricular ICD?
Does reducing potassium in the dialysis bath resolve hyperkalemia-induced T wave oversensing and inappropriate ICD shocks in a patient with a biventricular ICD?
Hyperkalemia during hemodialysis can cause T wave oversensing leading to inappropriate ICD shocks, which can be resolved by adjusting the dialysis bath potassium content.
Answers are grounded in this paper and cite their sources.
May support dialysate potassium adjustment to avert hyperkalemia-related ICD oversensing in dialysis patients; leaves open generalizability from this single case.
Inappropriate ICD shocks remain a common problem. Double counting of single ventricular events can occur with biventricular ICDs implanted before univentricular sensing was available. Often this is due to a tachyarrhythmia or loss of left ventricular capture. This report describes a patient who developed hyperkalemia during hemodialysis, received inappropriate ICD shocks and experienced loss of biventricular pacing due to T wave rather than QRS double counting. Oversensing was abolished by reducing the potassium content of the dialysis bath. This underscores the need for careful interpretation of saved electrograms to determine the cause for, and appropriate treatment of, device related problems.
No takes yet. Share an insight, caveat, or question.
Koul et al. (2004) studied this question.