Key result
Simultaneous intracardiac and intravascular ultrasound guidance for off-pump orthotopic aortic valved stent implantation was feasible in an animal model, achieving full valvular opening and closing in 60% of cases.
Why the study?
Does simultaneous intracardiac and intravascular ultrasound guidance allow adequate dimension assessment and deployment monitoring for off-pump aortic valved stent implantation in a porcine model?
Does simultaneous intracardiac and intravascular ultrasound guidance allow adequate dimension assessment and deployment monitoring for off-pump aortic valved stent implantation in a porcine model?
Simultaneous intracardiac and intravascular ultrasound guidance is feasible for monitoring and dimension assessment during off-pump aortic valved stent implantation in a porcine model.
Answers are grounded in this paper and cite their sources.
Feasibility in porcine model leaves open human translation of ultrasound-guided off-pump aortic stenting; prospective trials needed.
PURPOSE: To validate the use of simultaneous intracardiac and intravascular ultrasound (IVUS) guidance for periprocedural dimension assessment, Residual-Coronary/Sinus-Stent Index (RCSSI) estimation, target site identification, deployment monitoring, and quality control of off-pump aortic Valved Stent implantation. METHODS: Five pigs (56+/-5 kg) underwent off-pump orthotopic aortic valve implantation using a custom-made self-expanding Valved Stent. Intracardiac ultrasound (AcuNav) was introduced via the right femoral vein. After left-sided thoracotomy, pursestring sutures were placed on the left ventricular apex. Following heparinization, a guidewire was inserted through the apex and advanced over the aortic valve under fluoroscopy. A wire-guided IVUS catheter transducer (6-F, 12.5-MHz) was inserted and the aortic target site identified. IVUS probe location was tracked with AcuNav, and measures of the aortic root were taken by both. After removal of the IVUS, the Valved Stent delivery system was introduced over the guidewire under fluoroscopy and AcuNav monitoring; the Valved Stent was deployed over the native valves. In vivo assessment included leaflet motion, planimetric valve orifice and RCSSI (stent to aortic wall distance/coronary diameter) calculations, coronary blood flow characteristics, transvalvular gradient, regurgitation, and paravalvular leaking in combination with continuous cardiac output measures. Macroscopic analysis was performed at necropsy. RESULTS: IVUS dimensions of the aortic root were equal to AcuNav and necropsy dimensions. Both tools showed good valvular function, with full valvular opening and closing in 3 of 5 valves. At necropsy, the 3 aortic Valved Stents were safely anchored. One Valved Stent was placed supra-annularly; 2 dislodged into the left ventricle because of size mismatch. One Valved Stent showed a moderate to severe paravalvular leak. CONCLUSIONS: Simultaneous intracardiac and intravascular ultrasound-guided off-pump orthotopic aortic Valved Stent implantation via left sided thoracotomy is feasible in an animal model. IVUS and intracardiac ultrasound allow adequate aortic dimension assessment and Valved Stent delivery monitoring, as well as postimplantation coronary flow evaluation.
No takes yet. Share an insight, caveat, or question.
Huber et al. (2004) studied Off-pump aortic valved stent implantation (n=5). Simultaneous intracardiac and intravascular ultrasound (IVUS) guidance was evaluated on Feasibility of ultrasound-guided off-pump orthotopic aortic Valved Stent implantation (full valvular opening and closing). Simultaneous intracardiac and intravascular ultrasound guidance for off-pump orthotopic aortic valved stent implantation was feasible in an animal model, achieving full valvular opening and closing in 60% of cases.