Key result
A newly developed transluminal valved stent demonstrated good hemodynamic function in vitro and in vivo in pigs, though positioning and anchoring remained difficult with some stents tilting or migrating.
Early preclinical development of a transluminal heart valve prosthesis demonstrated feasible hemodynamic function but highlighted challenges in positioning and anchoring.
Answers are grounded in this paper and cite their sources.
Animal data support hemodynamic feasibility of transluminal valved stents; leaves open anchoring and positioning challenges before clinical translation.
The aim of this study was to develop a heart valve prosthesis that could be transluminally inserted through the infrarenal aorta in pigs and the femoral artery in humans. Therefore, this valve prosthesis had to be foldable to a diameter of approximately 8 mm, self-expandable when released in the vascular system, and equipped with an anchoring mechanism. For implantation, a suitable catheter delivery system for insertion via the abdominal aorta in pigs also was designed. Furthermore, an operation method, including positioning of the catheter delivery system and deployment of the valved stent under fluoroscopic and echocardiographic control, was developed. A series of different prototypes of the valved stents and catheter delivery systems was produced to optimize design and handling. These prototypes were tested in vitro in a circulation model, and those showing satisfying hemodynamic properties were implanted in pigs as in vivo studies afterwards. The valved stents had a good hemodynamic function in vitro and in vivo with no more than a mild regurgitation or stenosis. Valve movements were satisfying, and the design proved to be generally feasible. However, positioning and anchoring were still difficult; some stents were tilted in the lumen or migrated after implantation.
No takes yet. Share an insight, caveat, or question.
Kuklinski et al. (2004) studied Aortic valve replacement. Transluminal implantation of a foldable, self-expandable valved stent was evaluated on Hemodynamic function, valve movements, positioning, and anchoring. A newly developed transluminal valved stent demonstrated good hemodynamic function in vitro and in vivo in pigs, though positioning and anchoring remained difficult with some stents tilting or migrating.
Synapse has enriched 10 related papers. Consider them for comparative context: