Key result
Implantation of a re-positionable aortic stent-valve in swine demonstrated good valve function and maintained diastolic pressure, whereas the stent-without-valve resulted in massive regurgitation.
Why the study?
Does a re-positionable stent-valve improve valve function compared to a stent-without-valve in a swine model?
Does a re-positionable stent-valve improve valve function compared to a stent-without-valve in a swine model?
A newly designed re-positionable aortic stent-valve is feasible and effective in a swine model, though complicated by early thrombus formation.
Answers are grounded in this paper and cite their sources.
Preclinical swine data warrant no clinical adoption; leaves open human translation pending thrombus mitigation and safety trials.
BACKGROUND: Placement of aortic stent-valves in the proper position is difficult due to the high velocity and oscillation of blood flow and anatomical characteristics. To address this problem, a re-positionable stent-valve was designed and examined in a preliminary animal experiment. METHODS: Protocols for this animal study were approved by the institutional Animal Care and Use Committee. An umbrella-shaped polyurethane valve was attached at the tip of newly devised re-positionable metallic stent. The prosthesis (re-positionable stent-valve) was implanted in 5 female pigs via the right carotid artery using a 10-F introducer (Group A). As a control, the stent-without-valve was implanted in 4 female pigs (Group B). Efficacy of the valve was assessed by aortography and aortic pressure measurement. Animals were sacrificed 1 hour after stent-valve implantation for macroscopic examination. RESULTS: The stent-valve was successfully implanted in all pigs. In Group A, aortography showed good function of the polyurethane valve. No significant change in diastolic pressure was noted during valve testing. In Group B, aortography and aortic pressure measurement showed massive aortic regurgitation following dysfunction of the native aortic leaflets. Post-mortem examination showed small amounts of thrombus inside the polyurethane umbrella in Group A. CONCLUSION: Placement of the re-positionable stent-valve seems feasible and effective. However, the problem of thrombus formation inside the umbrella valve should be solved in future studies.
No takes yet. Share an insight, caveat, or question.
Hashimoto et al. (2008) studied Aortic valve replacement (animal model) (n=9). Re-positionable aortic stent-valve vs. Stent-without-valve was evaluated on Efficacy assessed by aortography and aortic pressure measurement. Implantation of a re-positionable aortic stent-valve in swine demonstrated good valve function and maintained diastolic pressure, whereas the stent-without-valve resulted in massive regurgitation.
Synapse has enriched 10 related papers. Consider them for comparative context: