Key result
A novel pre-clinical porcine model using a composite valved conduit allowed for site-specific in-vivo hemodynamic assessment of aortic valvular prostheses without cardiopulmonary bypass, demonstrating no significant difference in cardiac output compared to the native valve.
Why the study?
Can a pre-clinical porcine model be developed for site-specific in-vivo hemodynamic assessment of aortic valvular prosthesis without cardiopulmonary bypass?
Can a pre-clinical porcine model be developed for site-specific in-vivo hemodynamic assessment of aortic valvular prosthesis without cardiopulmonary bypass?
Absolute Event Rate: 4.3% vs 4.5%
p-value: p=NS
A novel pre-clinical porcine model allows for in-vivo hemodynamic assessment of aortic valvular prostheses without the need for cardiopulmonary bypass.
Facilitates preclinical aortic valve testing without bypass in animals; leaves open translation to clinical prosthesis assessment.
The objective of this study was to develop a pre-clinical large animal model for the in vivo hemodynamic testing of prosthetic valves in the aortic position without the need for cardiopulmonary bypass. Ten male pigs were used. A composite valved conduit was constructed in the operating room by implanting a prosthetic valve between two separate pieces of vascular conduits, which bypassed the ascending aorta to the descending aorta. Prior to applying a side-biting clamp to the ascending aorta for proximal grafting to the aortic anastomosis, an aorta to femoral artery shunt was placed just proximally to this clamp. The heart rate, cardiac output, Vmax, transvalvular pressure gradient, effective orifice area and incremental dobutamine stress response were assessed. A dose-dependent increase with dobutamine was seen in terms of cardiac output, Vmax, and the peak transvalvular pressure gradient both in the native and in the prosthetic valve. However, the increment was much steeper in the prosthetic valve. No significant differences in cardiac output were noted between the native and the prosthetic valves. The described pre-clinical porcine model was found suitable for site-specific in-vivo hemodynamic assessment of aortic valvular prosthesis without cardiopulmonary bypass.
No takes yet. Share an insight, caveat, or question.
Choo et al. (2004) studied Animal model for prosthetic valve testing (n=10). Composite valved conduit with mechanical aortic heart valve vs. Native aortic valve was evaluated on Cardiac output (L/min) (p=NS). A novel pre-clinical porcine model using a composite valved conduit allowed for site-specific in-vivo hemodynamic assessment of aortic valvular prostheses without cardiopulmonary bypass, demonstrating no significant difference in cardiac output compared to the native valve.