Key result
Bilateral septal pacing combined with coronary venous pacing narrowed QRS duration to 122 ± 10 ms compared to 156 ± 9 ms with conventional CRT.
Why the study?
QRS duration with conventional CRT frequently exceeds 130 ms, prompting evaluation of bilateral septal pacing combined with coronary venous pacing (BSP-CRT) for QRS narrowing.
Does bilateral septal pacing combined with coronary venous pacing improve QRS narrowing compared to conventional CRT in patients with QRS > 130 ms?
Observational (n=14)
Does bilateral septal pacing combined with coronary venous pacing improve QRS narrowing compared to conventional CRT in patients with QRS > 130 ms?
Answers are grounded in this paper and cite their sources.
Bilateral septal pacing combined with coronary venous pacing provides superior acute electrical synchronization compared to conventional CRT in patients with residual QRS > 130 ms.
BACKGROUND: Conventional right ventricular pacing combined with coronary venous pacing (CVP) is a mainstay for cardiac resynchronization therapy (CRT). However, QRS duration of conventional CRT may be frequently more than 130 ms. This study aimed to evaluate the effectiveness of QRS narrowing by bilateral septal pacing (BSP) in combination with CVP for CRT (BSP-CRT). METHODS: Fourteen patients with QRS > 130 ms of conventional CRT after failure of physiological conduction system pacing were enrolled. Electrophysiologic characteristics were compared among different modes of CRT during procedure. BSP which was defined as capture of both sides of interventricular septum manifested as shortened R wave peak time without a right bundle branch block QRS pattern. RESULTS: BSP-CRT were successfully achieved in 85.7% (12/14) patients. QRS duration at baseline was 185 ± 13 ms and significantly narrowed to 156 ± 9 ms during conventional CRT (n = 14, P < .001), to 143 ± 7 ms during left ventricular septal pacing (LVSP) in combination with CVP for CRT (LVSP-CRT) (n = 9, P < .001), and further to 122 ± 10 ms during BSP-CRT (n = 12, P < .001). Notably, among 7 patients in whom both LVSP and BSP were achieved, BSP-CRT outperformed LVSP-CRT at QRS narrowing by 16% (P < .001). At 3-month follow-up, left ventricular ejection fraction improved from 29 ± 6% to 41 ± 8% (P < .001). CONCLUSIONS: BSP-CRT resulted in superior acute electrical synchronization to conventional CRT and might be considered as an alternative to conventional CRT with QRS more than 130 ms.
No takes yet. Share an insight, caveat, or question.
Li et al. (2022) conducted an observational in QRS > 130 ms of conventional CRT after failure of physiological conduction system pacing (n=14). Bilateral septal pacing in combination with coronary venous pacing (BSP-CRT) vs. Conventional CRT was evaluated on QRS duration (p=<.001). Bilateral septal pacing combined with coronary venous pacing narrowed QRS duration to 122 ± 10 ms compared to 156 ± 9 ms with conventional CRT.