Key result
Reprogramming an ICD from DDD to DDIR terminates PMT and prevents further syncope.
Case Report (n=1)
This case highlights that pacemaker-mediated tachycardia can induce severe hemodynamic instability and syncope, which can be resolved by meticulous device evaluation and reprogramming.
Device interrogation in ICD patients with syncope can reveal PMT; confirms it as a reversible cause in severe LV dysfunction.
A 78-year-old male with a history of ischemic heart disease with prior myocardial infarction, severe left ventricular systolic dysfunction, and a device-related implantable cardioverter defibrillator (DR ICD) was admitted following a road traffic accident due to a sudden loss of control of the vehicle following a syncope. Initial investigations ruled out acute coronary syndrome and did not point to neurological causes. Device interrogation revealed the presence of pacemaker-mediated tachycardia (PMT), but since the episode trigger had been programmed off, it was not possible to determine when they occurred, how long they lasted, or their rate. However, atrial threshold testing promptly provoked a PMT. The induced PMT led to a tachycardia of 130 bpm, causing severe symptomatic hypotension with a mean blood pressure falling below 50 mmHg. This could be easily reproduced several times. Given the patient's severe left ventricular systolic dysfunction and the absence of any overt alternative explanation, the PMT likely triggered the syncope preceding the car accident. Management focused on reprogramming the DR ICD. The most critical adaptation was a mode switch from DDD to DDIR to prevent atrial tracking, effectively terminating PMT episodes. A prolongation of the postventricular atrial refractory period (PVARP) was not possible due to the very slow VA conduction. The follow-up revealed no other episodes of PMT and the patient no longer experiences episodes of presyncope or syncope. This report highlights the potential for PMTs to induce severe hemodynamic instability and syncope, underscoring the importance of meticulous device evaluation.
No takes yet. Share an insight, caveat, or question.
Grogg et al. (2026) conducted a case report in Pacemaker-mediated tachycardia (n=1). Device reprogramming (mode switch from DDD to DDIR) was evaluated. Reprogramming a dual-chamber ICD from DDD to DDIR mode effectively terminated pacemaker-mediated tachycardia and prevented further episodes of syncope in a 78-year-old male.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: