Key result
Clockwise neck rotation caused a 9-second ventricular asystolic pause and syncope due to bipolar ventricular lead insulation damage (impedance <250 Ω), resolved by adding a new ventricular lead.
Case Report (n=1)
No
In pacemaker patients presenting with unexplained recurrent syncope and normal resting device interrogations, prolonged ECG monitoring and provocative body movement maneuvers during interrogation should be performed to unmask intermittent lead insulation failure.
Answers are grounded in this paper and cite their sources.
Provocative maneuvers may unmask positional lead failure in unexplained syncope; hypothesis-generating and requires prospective validation before adoption.
Syncope following permanent pacemaker (PM) implantation is a nightmare for electrophysiologists. We describe a case of daily recurrent syncope in an 84-year-old man having a dual-chamber pacemaker implanted for complete atrio-ventricular block occurred 4 years before the admission to our department. He had a history of arterial hypertension, parossistic atrial fibrillation, chronic obstructive pulmonary disease, stage-III chronic renal failure, mild vascular cognitive impairment and glaucoma. The initial work-up including electrocardiogram (ECG), repeated PM interrogations, Holter electrocardiogram, blood pressure measurement in orthostatic position, complete blood count, serum glycaemia, electrolytes and thyroid function tests showed normal findings. Syncope occurred in lying position and during 90° left clockwise neck rotation and was associated to pallor, sweating, tonic-clonic seizures and transient self-limited loss of consciousness lasting a few seconds. Electroencephalogram was normal. During continuous ECG monitoring, the right rotation of the head determined a ventricular asystolic pause lasting 9 seconds associated with loss of consciousness. Restoration of sinus rhythm was observed after bringing back the head in axis. The PM interrogation, performed during pacing failure, recorded low impedance of bipolar ventricular lead, suggesting a damage in lead insulation. It is likely that lead movements during clockwise neck rotation produced an intermittent short circuit that prevented sufficient energy delivery to the myocardium with a consequence of sudden loss of capture.
No takes yet. Share an insight, caveat, or question.
Mazza et al. (2012) conducted a case report in Recurrent syncope in a patient with a permanent dual-chamber pacemaker implanted for complete atrioventricular block (n=1). Pacemaker replacement with a 3-chamber system and implantation of a new bipolar steroid-eluting active-fixation right ventricular lead was evaluated on Resolution of pacing failure and restoration of stable pacing threshold and impedance. Clockwise neck rotation caused a 9-second ventricular asystolic pause and syncope due to bipolar ventricular lead insulation damage (impedance <250 Ω), resolved by adding a new ventricular lead.
Synapse has enriched 10 related papers. Consider them for comparative context: