Key result
Lead-header malapposition caused late permanent pacemaker malfunction and recurrent syncope 14 months after initial device implantation.
Why the study?
Permanent pacemaker malfunction due to lead-header malapposition is extremely rare and has not previously been reported presenting late after implantation.
Case Report (n=1)
Lead-header malapposition can present as a rare, late complication of permanent pacemaker implantation causing recurrent syncope.
Consider lead-header issues in unexplained late syncope after pacing; leaves open prospective incidence studies.
Permanent pacemaker (PPM) malfunction due to electrical connection problems such as a loose set screw or lead-header malapposition is extremely rare. We present a patient with complete heart block (CHB) who had PPM malfunction and recurrent syncope, late (14 months) after initial implantation, which was caused by the ventricular lead pin disengagement from the header resulting in oversensing due to noise, pacing inhibition and recurrent syncope. PPM due to lead-header malapposition this late after device implantation has previously not been reported.
No takes yet. Share an insight, caveat, or question.
Mazzone et al. (2020) conducted a case report in Complete heart block (n=1). Lead-header malapposition was evaluated on Pacemaker malfunction and recurrent syncope. Lead-header malapposition caused late permanent pacemaker malfunction and recurrent syncope 14 months after initial device implantation.
Synapse has enriched 5 closely related papers on similar clinical questions. Consider them for comparative context: