Key result
Extraction of a retained, inactive pacemaker lead completely resolved recurrent ventricular tachycardia and syncope caused by a redundant lead loop in the right ventricle.
Answers are grounded in this paper and cite their sources.
Retained inactive pacemaker leads can cause mechanical irritation leading to ventricular tachycardia, highlighting the importance of complete lead extraction when feasible.
Our patient had recurrent syncope due to ventricular tachycardia (VT) after one year of VVI Pacemaker implantation. He had pacemaker pocket infection for which new pacemaker was implanted on opposite side but old lead was not explanted completely. Flouroscopy showed redundant loop of old lead in right ventricular inflow which was snared out subsequently. He never had syncope or VT after that.
No takes yet. Share an insight, caveat, or question.
Datta et al. (2011) studied this question. Extraction of a retained, inactive pacemaker lead completely resolved recurrent ventricular tachycardia and syncope caused by a redundant lead loop in the right ventricle.