The reconfirmation algorithm in third-generation ICDs can lead to inadequate shocks when VT spontaneously converts to AF with a slow ventricular response.
Alerts clinicians to possible inadequate shocks during VT-to-AF conversion; leaves open whether reconfirmation algorithms require modification in third-generation ICDs.
We report on inadequate shocks delivered by an ICD for VT converting spontaneously to AF with slow ventricular response. This represents a correct behavior of the reconfirmation algorithm, as the detection of intervals longer than the fibrillation detection interval and shorter than the bradycardia escape interval is required to inhibit the shock.
No takes yet. Share an insight, caveat, or question.
MATTKE et al. (1997) studied this question.